FDA Adverse Event Injury Summary report: N

HAHN TAPERED IMPLANT Ø3.5 X 10 MM

MDR report key: 21540770 · Received March 6, 2025

Report

Report Number
3011649314-2025-00180
Event Type
Injury
Date Received
March 6, 2025
Date of Event
November 21, 2024
Report Date
May 9, 2025
Manufacturer
PRISMATIK DENTALCRAFT, INC.
Product Code
DZE
PMA / PMN Number
K143353
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE COMPLAINT DEVICE HAS NOT BEEN RETURNED FOR INVESTIGATION. COMPLAINT HISTORY AND PRODUCT HISTORY RECORDS WERE REVIEWED, AND DOCUMENTATION INDICATED THE PRODUCT MET RELEASE CRITERIA. SHOULD THE DEVICE BE RETURNED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. FILE LINKED TO SUBMISSION NAME: 3011649314-2025-00179. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: COMP-(B)(4).

Additional Manufacturer Narrative · 0

DEVICE EVALUATION HAS BEEN COMPLETED; FOLLOWING IS THE DEVICE ANALYSIS CONCLUSION: DHR RESULTS THE DHR WAS REVIEWED FOR HAHN TAPERED IMPLANT LOT# 6201827 AND THERE WAS NO EVIDENCE DISCOVERED TO INDICATE THAT A PRODUCT DEFECT OR NON-CONFORMITY CONTRIBUTED TO THE ISSUE. THE PART MET ALL THE CRITERIA CALLED FOR IN THE PRODUCTION ROUTER. STOCK PRODUCT REVIEWED RESULTS THE STOCK PRODUCT FOR HAHN TAPERED IMPLANT LOT# 6201827 WAS REVIEWED AND FOUND TO BE IN STOCK, BUT IT IS NOT APPLICABLE FOR PHYSICAL EVALUATION SINCE THE ISSUE IS CUSTOMER RELATED. INVESTIGATION METHODS/RESULTS THE DEVICE WAS RETURNED BUT NOT IN THE ORIGINAL PACKAGE. THE IMPLANT WAS VERIFIED TO BE A HAHN TAPERED IMPLANT Ø3.5 X 10 MM (70-1154-IMP0005) USING THE RADIOGRAPHIC TEMPLATE (PK-209-062515). THERE WAS NO DEFECT OR NON-CONFORMITY OBSERVED AND THE THREADS WERE INTACT. MATTER WAS OBSERVED IN THE TREADING OF THE IMPLANT. THE COMPLAINT IS VERIFIED BASED ON THE RETURNED PART BUT CANNOT CONFIRM THE FAILURE MODE. THERE WAS NO EVIDENCE FOUND THAT INDICATED THAT THE REPORTED ISSUE WAS CAUSED BY THE DEVICE ITSELF. ROOT CAUSE "FAILURE TO OSSEOINTEGRATE" IS A COMMON COMPLAINT IN REGARD TO IMPLANT FAILURE. THIS OCCURS WHEN THE PATIENT'S BONE DOES NOT INTEGRATE WITH THE IMPLANT SURFACE. THE POSSIBLE RESPONSES TO THIS COMPLAINT COULD BE ATTRIBUTED TO VARIOUS CAUSES. ALTHOUGH THE ROOT CAUSE FOR FAILURE TO OSSEOINTEGRATE IS INCONCLUSIVE AND SPECIFIC TO EACH CASE, PROBABLE CAUSES COULD BE DUE TO INSUFFICIENT BONE OR POOR BONE QUALITY; EITHER THE BONE WAS TOO SOFT, OR THE OPERATOR ERRED IN CREATING AN OSTEOTOMY BIGGER THAN THE SIZE OF THE IMPLANT DIAMETER. PREMATURE LOADING, PATIENT'S HEALTH, PERI-IMPLANTITIS, SMOKING, AND LACK OF ORAL HYGIENE MAY ALSO BE CONTRIBUTING FACTORS. IFU-570 REV 4 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN THE CONTRAINDICATIONS SECTION: "HAHN TAPERED IMPLANTS SHOULD NOT BE PLACED IN PATIENTS DISCOVERED TO BE MEDICALLY UNFIT FOR THE INTENDED TREATMENT. PRIOR TO CLINICAL INTERVENTION, PROSPECTIVE PATIENTS MUST BE THOROUGHLY EVALUATED FOR ALL KNOWN RISK FACTORS AND CONDITIONS RELATED TO ORAL SURGICAL PROCEDURES AND SUBSEQUENT HEALING. CONTRAINDICATIONS INCLUDE BUT ARE NOT LIMITED TO: VASCULAR CONDITIONS, UNCONTROLLED DIABETES, CLOTTING DISORDERS, ANTICOAGULANT THERAPY, METABOLIC BONE DISEASE, CHEMOTHERAPY OR RADIATION THERAPY, CHRONIC PERIODONTAL INFLAMMATION, INSUFFICIENT SOFT TISSUE COVERAGE, METABOLIC OR SYSTEMIC DISORDERS ASSOCIATED WITH WOUND AND/OR BONE HEALING, USE OF PHARMACEUTICALS THAT INHIBIT OR ALTER NATURAL BONE REMODELING, ANY DISORDERS WHICH INHIBIT A PATIENT'S ABILITY TO MAINTAIN ADEQUATE DAILY ORAL HYGIENE, UNCONTROLLED PARAFUNCTIONAL HABITS, INSUFFICIENT HEIGHT AND/OR WIDTH OF BONE, AND INSUFFICIENT INTERARCH SPACE." IFU-570 REV 4 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN THE SURGICAL PROCEDURES UNDER THE PRECAUTION SECTION: "MINIMIZING TISSUE DAMAGE IS CRUCIAL TO SUCCESSFUL IMPLANT OSSEOINTEGRATION. IN PARTICULAR, CARE SHOULD BE TAKEN TO ELIMINATE SOURCES OF INFECTION, CONTAMINANTS, SURGICAL AND THERMAL TRAUMA. RISK OF OSSEOINTEGRATION FAILURE INCREASES AS TISSUE TRAUMA INCREASES. FOR BEST RESULTS, PLEASE OBSERVE THE FOLLOWING PRECAUTIONS: ALL DRILLING PROCEDURES SHOULD BE PERFORMED AT 2000 RPM OR LESS UNDER CONTINUAL, COPIOUS IRRIGATION. ALL SURGICAL INSTRUMENTS USED MUST BE IN GOOD CONDITION AND SHOULD BE USED CAREFULLY TO AVOID DAMAGE TO IMPLANTS OR OTHER COMPONENTS. IMPLANTS SHOULD BE PLACED WITH SUFFICIENT STABILITY; HOWEVER, EXCESSIVE INSERTION TORQUE MAY RESULT IN IMPLANT FRACTURE, OR FRACTURE OR NECROSIS OF THE IMPLANT SITE. THE PROPER SURGICAL PROTOCOL SHOULD BE STRICTLY ADHERED TO." IFU-570 REV 4 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN THE WARNING SECTION: "ABSOLUTE SUCCESS CANNOT BE GUARANTEED. FACTORS SUCH AS INFECTION, DISEASE, AND INADEQUATE BONE QUALITY AND/OR QUANTITY CAN RESULT IN OSSEOINTEGRATION FAILURES FOLLOWING SURGERY OR INITIAL OSSEOINTEGRATION." IFU-570 REV 4 (HAHN TAPERED IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN THE WARNING SECTION: "THE IMPLANT SITE SHOULD BE INSPECTED FOR ADEQUATE BONE BY RADIOGRAPHS, PALPATIONS AND VISUAL EXAMINATION. DETERMINE THE LOCATION OF NERVES AND OTHER VITAL STRUCTURES AND THEIR PROXIMITY TO THE IMPLANT SITE BEFORE ANY DRILLING TO AVOID POTENTIAL INJURY, SUCH AS PERMANENT NUMBNESS TO THE LOWER LIP AND CHIN." THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Description of Event or Problem · 0

A HEALTHCARE PROFESSIONAL REPORTED THAT A HAHN TAPERED IMPLANT FAILED. THE PATIENT'S BONE QUALITY IS TYPE LL OR III. THE PATIENT PRESENTED ON (B)(6) 2024 FOR A PRIMARY PROCEDURE ON TEETH 4 AND 5. AT THE SECOND STAGE OF SURGERY, ON (B)(6) 2024 THE PROVIDER DETERMINED THAT THE IMPLANTS FAILED TO OSSEOINTEGRATE, AND THE IMPLANTS WERE REMOVED. THE SITES WERE GRAFTED, AND THE IMPLANTS WERE NOT REPLACED. NO PERMANENT INJURY WAS REPORTED. PER THE REPORTED INFORMATION, THE PATIENT HAS NO PREEXISTING MEDICAL HISTORY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
536802 HAHN TAPERED IMPLANT Ø3.5 X 10 MM HAHN TAPERED IMPLANT SYSTEM DZE PRISMATIK DENTALCRAFT, INC. 70-1154-IMP0005 6201827

Patients

Seq Age Sex Outcome Treatment
1 58 YR Male Required Intervention