FDA Adverse Event Malfunction Summary report: N

BALLARD CLOSED SUCTION SYSTEM

MDR report key: 21540215 · Received March 6, 2025

Report

Report Number
8030647-2025-00022
Event Type
Malfunction
Date Received
March 6, 2025
Date of Event
February 6, 2025
Report Date
June 10, 2025
Manufacturer
AVANOS MEDICAL INC.
Product Code
BSY
UDI-DI
00609038938240
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
117

Narratives

Additional Manufacturer Narrative · 0

CORRECTION: D1. ADDITIONAL INFORMATION: H6 THE PRODUCT INVOLVED IN THE REPORT HAS NOT BEEN RETURNED FOR EVALUATION; HOWEVER, THE REPORTER PROVIDED PHOTOGRAPHIC /VIDEO EVIDENCE WHICH DEPICTS THE REPORTED EVENT: BROKEN Y-ADAPTOR; THE COMPLAINT IS CONSIDERED TO BE CONFIRMED. INVESTIGATION FOR THIS FAILURE MODE WAS PREVIOUSLY PERFORMED; AS THE ISSUE IS A KNOWN FAILURE; THE ROOT CAUSE WAS DETERMINED TO BE MANUFACTURING RELATED. PROCESSES ARE IN PLACE TO PREVENT FUTURE REOCCURRENCE OF THE ISSUE. THE DEVICE HISTORY RECORD FOR LOT 4301085 WAS REVIEWED AND THE PRODUCT WAS PRODUCED ACCORDING TO PRODUCT SPECIFICATIONS. ALL INFORMATION REASONABLY KNOWN AS OF 10 JUN 2025 HAS BEEN INCLUDED IN THIS HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED, A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. THE INFORMATION PROVIDED BY SUN MED HOLDINGS LLC. REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO SUN MED HOLDINGS LLC. SUN MED HOLDINGS LLC. HAS NO INDEPENDENT KNOWLEDGE OF THE EVENT REPORTED BUT IS RELAYING THE INFORMATION THAT WAS PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE SUN MED HOLDINGS COMPLAINT DATABASE AND IDENTIFIED AS (B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO 21CFR803, IN COMPLIANCE WITH THE MEDICAL DEVICE REPORTING REQUIREMENT AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT AN SUN MED HOLDINGS LLC. PRODUCT IS DEFECTIVE OR CAUSED SERIOUS INJURY.

Additional Manufacturer Narrative · 0

THE PRODUCT INVOLVED IN THE REPORT HAS NOT BEEN RETURNED FOR EVALUATION. A REVIEW OF THE DEVICE HISTORY RECORD IS IN-PROGRESS. ALL INFORMATION REASONABLY KNOWN AS OF 06 MAR 2025 HAS BEEN INCLUDED IN THIS HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED, A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. THE INFORMATION PROVIDED BY SUN MED HOLDINGS LLC. REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO SUN MED HOLDINGS LLC. SUN MED HOLDINGS LLC. HAS NO INDEPENDENT KNOWLEDGE OF THE EVENT REPORTED BUT IS RELAYING THE INFORMATION THAT WAS PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE SUN MED HOLDINGS COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT-(B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO 21CFR803, IN COMPLIANCE WITH THE MEDICAL DEVICE REPORTING REQUIREMENT AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT AN SUN MED HOLDINGS LLC. PRODUCT IS DEFECTIVE OR CAUSED SERIOUS INJURY.

Description of Event or Problem · 0

IT WAS REPORTED, THE BALLARD IN-LINE SUCTION SNAPPED OFF WITH THE SPIKE BEING STUCK IN THE OPENING OF THE 3.0 ENDOTRACHEAL TUBE (ETT), THIS MADE IT IMPOSSIBLE TO BAG THE PATIENT. THE RESPIRATORY THERAPIST (RT) CUT THE ETT AND INSERT A NEW BALLARD IN ORDER TO BAG THE PATIENT; THERE WAS NO PATIENT HARM.

Description of Event or Problem · 0

IT WAS REPORTED, THE BALLARD IN-LINE SUCTION SNAPPED OFF WITH THE SPIKE BEING STUCK IN THE OPENING OF THE 3.0 ENDOTRACHEAL TUBE (ETT), THIS MADE IT IMPOSSIBLE TO BAG THE PATIENT. THE RESPIRATORY THERAPIST (RT) CUT THE ETT AND INSERT A NEW BALLARD IN ORDER TO BAG THE PATIENT; THERE WAS NO PATIENT HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
436329 BALLARD CLOSED SUCTION SYSTEM BALLARD CLOSED SUCTION CATHETER, Y-ADAPTER, NEONATAL/PEDIATRIC BSY AVANOS MEDICAL INC. 93824 4301085 00609038938240

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown