SOLYX BLUE
Report
- Report Number
- 2124215-2025-12579
- Event Type
- Malfunction
- Date Received
- March 6, 2025
- Date of Event
- February 11, 2025
- Report Date
- September 2, 2025
- Manufacturer
- BOSTON SCIENTIFIC CORPORATION
- Product Code
- PAH
- UDI-DI
- 08714729961901
- PMA / PMN Number
- K081275
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
BLOCK H6: DEVICE CODE A150205 CAPTURES THE REPORTABLE EVENT OF DIFFICULT TO ADVANCE.
BLOCK H11: UPON RECEIPT AT OUR QUALITY ASSURANCE LABORATORY, THE SOLYX BLUE DEVICE UNDERWENT A THOROUGH ANALYSIS. DURING THE PRODUCT ANALYSIS, IT WAS OBSERVED THAT THE MESH CARRIER WAS DETACHED, AND THE MESH WAS STRETCHED, CONFIRMING THE COMPLAINT. BASED ON THE INFORMATION AVAILABLE AND ANALYSIS RESULTS, GIVEN THAT MESH PLACEMENT IS LARGELY A BLIND PROCEDURE IN WHICH TOUGH MUSCLE, SOME BONE, AND OCCASIONALLY TENDONS MAY BE ENCOUNTERED, CREATING RESISTANCE DURING PLACEMENT--CONTRIBUTED TO THE ISSUE, A CONCLUSION CODE OF ADVERSE EVENT RELATED TO PROCEDURE WAS ASSIGNED TO THIS INVESTIGATION SINCE THE ADVERSE EVENTS OCCURRED DURING THE PROCEDURE AND THE DEVICE HAD NO INFLUENCE ON EVENT. BLOCK H6: DEVICE CODE A150205 CAPTURES THE REPORTABLE EVENT OF DIFFICULT TO ADVANCE.
IT WAS REPORTED THAT, DURING A PROCEDURE USING A SOLYX BLUE DEVICE, THE DEVICE HAD DIFFICULTY ADVANCING. THE PROCEDURE WAS COMPLETED USING ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
IT WAS REPORTED THAT, DURING A PROCEDURE USING A SOLYX BLUE DEVICE, THE DEVICE HAD DIFFICULTY ADVANCING. THE PROCEDURE WAS COMPLETED USING ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 487397 | SOLYX BLUE | MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR STRESS URINARY INCONTINENCE, FEMA | PAH | BOSTON SCIENTIFIC CORPORATION | M0068507010 | 0034957500 | 08714729961901 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | Unknown |