FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 2153733
·
Received July 7, 2011
Report
- Report Number
- 2531779-2011-04767
- Event Type
- Malfunction
- Date Received
- July 7, 2011
- Report Date
- June 14, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MT, US
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE PUMP HAS BEEN RETURNED TO ANIMAS FOR EVALUATION. THE FOLLOWING WAS FOUND WITH THE SUBJECT PUMP: ALL KEYPAD BUTTONS WERE RESPONDING INTERMITTENTLY. PRODUCT ANALYSIS REMOVED THE KEYPAD AND FOUND ADHESIVE UNDER THE CONTACTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
Description of Event or Problem · 1
THE PATIENT CLAIMED THAT THE UP/DOWN/OK BUTTONS REQUIRED SEVERAL PRESSES TO ACTIVATE THE DESIRED PUMP FUNCTIONS. THE KEYPAD IS REPORTEDLY INTACT. THERE WAS NO ADVERSE EVENT ASSOCIATE WITH THIS COMPLAINT. THE PUMP WAS REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM 2020 INSULIN INFUSION PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 18 YR |