HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1423500-2011-08786
- Event Type
- Malfunction
- Date Received
- July 7, 2011
- Date of Event
- June 15, 2011
- Report Date
- June 15, 2011
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE HAD BEEN DISCARDED AND THE LOT NUMBER WAS UNKNOWN; THEREFORE, A BATCH REVIEW COULD NOT BE PERFORMED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE A FOLLOW-UP REPORT WILL BE FILED.
(B)(4). THIS COMPLAINT FOR A CHECK PATIENT LINE ALARM WAS NOT CONFIRMED IN THE LAB. A SAMPLE EVALUATION WAS NOT PERFORMED DUE TO UNAVAILABLE SAMPLE. A BATCH REVIEW WAS NOT PERFORMED DUE TO UNKNOWN LOT NUMBER. A ROOT CAUSE FOR AN AIR BUBBLES IN PATIENT LINE WAS NOT IDENTIFIED DUE TO INSUFFICIENT INFORMATION. THE CAUSE WAS UNDETERMINED. RENAL QUALITY ENGINEERING, ALONG WITH PLANT QUALITY AND MANUFACTURING PERSONNEL, WILL CONTINUE TO MONITOR THIS PRODUCT LINE FOR TRENDS AND WILL TAKE CORRECTIVE/PREVENTIVE ACTION AS APPROPRIATE.
THE CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER (TSC) REGARDING A CHECK PATIENT LINE ALARM WHICH OCCURRED ON THE HOMECHOICE (HC) DURING INITIAL DRAIN. THE DRAIN VOLUME EQUALED 22ML. THE HP STATED THAT THERE WERE A LOT OF AIR BUBBLES IN HIS PATIENT LINE AND HE NEEDED TO START OVER WITH ALL NEW SUPPLIES. THE BAXTER TECHNICAL SERVICE REPRESENTATIVE (TSR) THEN ASSISTED THE HP TO CYCLE POWER OFF/ON, END THE THERAPY AND START OVER WITH ALL NEW SUPPLIES. THERE WAS PATIENT INVOLVEMENT BUT NO REPORT OF MEDICAL INTERVENTION OR PATIENT INJURY. PRODUCT SURVEILLANCE CONTACTED THE HOME PATIENT (HP) ON (B)(6) 2011 REGARDING THE CHECK PATIENT LINE ALARM. THE HOME PATIENT (HP) STATED THAT THE ISSUE WAS RESOLVED BY ENDING THERAPY AND CONTACTING THE NURSE IN THE MORNING. THE HP DID ONE MANUAL BAG IN THE MORNING AND CONTINUED THERAPY AS USUAL IN THE EVENING. HOWEVER, THE CAUSE OF THE ALARM AND THE AIR IN THE TUBING REMAINED UNKNOWN. THE HP VERIFIED THAT THERE WERE NO LOOSE CONNECTIONS, DISCONNECTIONS OR DEFECTS ON THE SUPPLIES. PER HP, HE DID NOT RECEIVE ANY INJURY OR MEDICAL INTERVENTION AS A RESULT OF THIS INCIDENT. THE HP STATED THAT HE IS DOING FINE AND CONTINUING THERAPY WITHOUT ISSUES. THE HP STATED THAT HE HAD DISCARDED THE SUPPLIES AFTER THERAPY, AND DID NOT KNOW THE LOT NUMBERS. NO ALLEGATIONS WERE MADE AGAINST ANY OF THE HP'S DIALYSIS PRODUCTS. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 51 YR | HOMECHOICE |