FDA Adverse Event Malfunction Summary report: N

FLEXCATH STEERABLE SHEATH

MDR report key: 2153714 · Received July 7, 2011

Report

Report Number
3002648230-2011-00074
Event Type
Malfunction
Date Received
July 7, 2011
Date of Event
June 8, 2011
Report Date
June 8, 2011
Manufacturer
MEDTRONIC CRYOCATH LP
Product Code
DRA
PMA / PMN Number
K081049
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HAS NOT YET BEEN RETURNED FOR EVALUATION. RESULTS OF EVALUATION OF RETURNED DEVICE WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED AND WAS VISUALLY INSPECTED AND FUNCTIONALLY TESTED. THE REPORTED ISSUE WAS CONFIRMED THROUGH TESTING. DISSECTION SHOWED A LEAKING HEMOSTATIC VALVE. A FIELD ACTION WAS INITIATED IN (B)(6) 2011 TO COMMUNICATE TO PHYSICIANS AND REGULATORY AUTHORITIES THIS POTENTIAL LEAK IN HEMOSTATIC VALVE OF MEDTRONIC CRYOCATH FLEXCATH 12 STEERABLE SHEATHS.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

A CRYOABLATION PROCEDURE WAS PERFORMED IN (B)(4). AFTER THE FIRST ABLATION, IT WAS NOTICED THAT THERE WAS A LEAK FROM THE VALVE OF THE FLEXCATH STEERABLE SHEATH (CATHETER INTRODUCER) AND AIR BUBBLES WERE SEEN IN THE TUBING OF THE SIDE PORT. THERE WAS NO PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLEXCATH STEERABLE SHEATH CATHETER, STEERABLE DRA MEDTRONIC CRYOCATH LP 3FC12 61557

Patients

Seq Age Sex Outcome Treatment
1 00066 YR