VITALITY
Report
- Report Number
- 2124215-2011-07799
- Event Type
- Malfunction
- Date Received
- July 7, 2011
- Date of Event
- April 7, 2011
- Report Date
- April 7, 2011
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, DEVICE INTERROGATION REVEALED THAT THE DEVICE BATTERY STATUS WAS ELECTIVE REPLACEMENT INDICATOR (ERI), WITH A BATTERY VOLTAGE OF 2.397 VOLTS. A LONGEVITY CALCULATION CONFIRMED THAT THE DEVICE DID NOT MEET LONGEVITY ESTIMATES PROVIDED IN DEVICE LABELING. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. FINAL ANALYSIS CONCLUDED THAT THE PREMATURE BATTERY DEPLETION WAS DUE TO TWO COMPROMISED LOW-VOLTAGE CAPACITORS, WHICH RESULTED IN A HIGH CURRENT CONDITION.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE WAS ELECTIVELY EXPLANTED. TO DATE, NO ADVERSE PATIENT EFFECTS WERE REPORTED WITH THIS CLINICAL OBSERVATION. THE DEVICE WAS RETURNED TO ALLOW FOR RELIABILITY ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITALITY | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | T180 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR | 4086| T180| 0184| E110 |