ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2011-04761
- Event Type
- Malfunction
- Date Received
- July 7, 2011
- Report Date
- June 13, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.
(B)(4). DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP'S ALARM HISTORY INDICATED MULTIPLE "REPLACE BATTERY" ALARMS HAD OCCURRED. THE PUMP POWERS ON NORMALLY WITH NO ALARMS. THERE WAS NO DAMAGE FOUND TO THE BATTERY CAP OR COMPARTMENT. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO POWER ISSUES OR ALARMS OCCURRING. THE PUMP ELECTRICAL DRAWS WERE TESTED AND FOUND TO BE OUT OF SPECIFICATION. A CASE SEAL LEAK WAS FOUND DURING LEAK TESTING, AND THERE WAS EVIDENCE OF INTERNAL MOISTURE DAMAGE FOUND. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THERE IS NO ADVERSE EVENT ASSOCIATED THIS COMPLAINT. THE PATIENT REPORTED THAT THE PUMP REQUIRES FREQUENT BATTERY CHANGES. HE STATED THAT THE BATTERY ONLY LASTS A COUPLE OF DAYS. HE REPORTED THAT THE BATTERY AND PUMP WERE HOT. HE TOOK BATTERY OUT AND LET THE PUMP COOL. HE INSERTED A NEW BATTERY AND THE PUMP WOULD NOT TURN ON. THE PATIENT ALSO STATED THAT OVERNIGHT THE PUMP BEEPED AND HE SAW THE VERIFY SCREEN REQUIRING FULL REWIND. THE PATIENT REPORTED WATER DAMAGE IN TOP LEFT OF THE SCREEN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 16 YR |