FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2153660 · Received July 7, 2011

Report

Report Number
2531779-2011-04761
Event Type
Malfunction
Date Received
July 7, 2011
Report Date
June 13, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP'S ALARM HISTORY INDICATED MULTIPLE "REPLACE BATTERY" ALARMS HAD OCCURRED. THE PUMP POWERS ON NORMALLY WITH NO ALARMS. THERE WAS NO DAMAGE FOUND TO THE BATTERY CAP OR COMPARTMENT. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO POWER ISSUES OR ALARMS OCCURRING. THE PUMP ELECTRICAL DRAWS WERE TESTED AND FOUND TO BE OUT OF SPECIFICATION. A CASE SEAL LEAK WAS FOUND DURING LEAK TESTING, AND THERE WAS EVIDENCE OF INTERNAL MOISTURE DAMAGE FOUND. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THERE IS NO ADVERSE EVENT ASSOCIATED THIS COMPLAINT. THE PATIENT REPORTED THAT THE PUMP REQUIRES FREQUENT BATTERY CHANGES. HE STATED THAT THE BATTERY ONLY LASTS A COUPLE OF DAYS. HE REPORTED THAT THE BATTERY AND PUMP WERE HOT. HE TOOK BATTERY OUT AND LET THE PUMP COOL. HE INSERTED A NEW BATTERY AND THE PUMP WOULD NOT TURN ON. THE PATIENT ALSO STATED THAT OVERNIGHT THE PUMP BEEPED AND HE SAW THE VERIFY SCREEN REQUIRING FULL REWIND. THE PATIENT REPORTED WATER DAMAGE IN TOP LEFT OF THE SCREEN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 16 YR