FDA Adverse Event
Injury
Summary report: N
VITALITY
MDR report key: 2153656
·
Received July 7, 2011
Report
- Report Number
- 2124215-2011-06408
- Event Type
- Injury
- Date Received
- July 7, 2011
- Date of Event
- April 6, 2011
- Report Date
- April 6, 2011
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Additional Manufacturer Narrative · 1
TO DATE, INFORMATION SUGGESTS THAT THIS ICD WAS REMOVED FROM SERVICE, BUT HAS NOT YET BEEN RETURNED TO BOSTON SCIENTIFIC. AS NEW INFORMATION WOULD BECOME AVAILABLE, THIS REPORT WILL BE FURTHER EVALUATED AND UPDATED.
Description of Event or Problem · 1
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD), WHICH IS PART OF THE MID-LIFE DISPLAY OF REPLACEMENT INDICATORS PRODUCT UPDATE CLASSIFIED AS A PRODUCT ADVISORY AND INITIALLY COMMUNICATED ON (B)(6) 2007, REACHED ELECTIVE REPLACEMENT INDICATOR (ERI) DUE TO EXTENDED CHARGE TIME MEASUREMENT WITH A MONITORING VOLTAGE MEASUREMENT OF 2.55 VOLTS. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS ASSOCIATED WITH THIS CLINICAL OBSERVATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITALITY | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | T125 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR | Required Intervention | 0157| 1861| T125| E110| 4087 |