CONTAK RENEWAL
Report
- Report Number
- 2124215-2011-06320
- Event Type
- Malfunction
- Date Received
- July 7, 2011
- Date of Event
- April 8, 2011
- Report Date
- April 13, 2011
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- NIK
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SD
- Reporter Occupation
- PHYSICIAN
Narratives
UPON COMPLETION OF ANALYSIS, THIS REPORT WILL BE UPDATED.
UPON RECEIPT AT OUR POST-MARKET QUALITY ASSURANCE LABORATORY, A LONGEVITY CALCULATION CONFIRMED THAT THE DEVICE DID NOT MEET LONGEVITY ESTIMATES PROVIDED IN DEVICE LABELING. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. FINAL ANALYSIS CONCLUDED THAT THE PREMATURE BATTERY DEPLETION WAS DUE TO LOW-VOLTAGE CAPACITOR DEGRADATION, WHICH RESULTED IN A HIGH CURRENT CONDITION.
BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS RETURNED FOR DISPOSAL AFTER EXPLANT. NO PRODUCT PERFORMANCE ALLEGATIONS OR ADVERSE PATIENT EFFECTS HAVE BEEN REPORTED WITH REGARD TO THIS DEVICE. ROUTINE INITIAL RETURNED DEVICE TESTING INDICATED THAT DETAILED ANALYSIS WAS REQUIRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONTAK RENEWAL | IMPLANTABLE CHF GENERATOR | NIK | GUIDANT CRM CLONMEL IRELAND | H170 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 82 YR | 0185| 4543| H170| 4087 |