FDA Adverse Event Malfunction Summary report: N

VITALITY 2

MDR report key: 2153562 · Received July 7, 2011

Report

Report Number
2124215-2011-06427
Event Type
Malfunction
Date Received
July 7, 2011
Date of Event
December 14, 2010
Report Date
April 7, 2011
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Removal / Correction Number
Z-0047-08 TO Z-0053-08
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED ELECTIVE REPLACEMENT INDICATOR (ERI) AFTER EXPERIENCING TWO CONSECUTIVE CHARGE TIMES GREATER THAN THE EXTENDED MID-LIFE ERI CHARGE TIME LIMIT. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, ERI WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE. THIS DEVICE IS INCLUDED IN THE MID-LIFE DISPLAY OF REPLACEMENT INDICATORS ((B)(6), 2007) ADVISORY POPULATION.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) HAD A VOLTAGE OF 2.73 AND CHARGE TIMES OF 18.4 SECONDS. THERE WAS INQUIRY OF THE CHARGE TIMES AND TECHNICAL SERVICES DISCUSSED DEVICE BEHAVIOR. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS DEVICE IS INCLUDED IN THE MID-LIFE DISPLAY OF REPLACEMENT INDICATORS ADVISORY. THE DEVICE REMAINED IN SERVICE AND WAS LATER EXPLANTED FOR NORMAL BATTERY DEPLETION. INITIAL ANALYSIS REVEALED A DEVICE ANOMALY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY 2 IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T177

Patients

Seq Age Sex Outcome Treatment
1 65 YR T177| 0184