FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 2153518
·
Received June 21, 2011
Report
- Report Number
- 1720753-2011-08369
- Event Type
- Malfunction
- Date Received
- June 21, 2011
- Date of Event
- June 14, 2011
- Report Date
- June 21, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE SURGE SUPPRESSOR AND THE INTELLIGENT SHUT-DOWN PRINTED CIRCUIT BOARDS WERE REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT BOOT UP DURING A CASE AND MADE A LOUD BEEPING NOISE. NO PT INJURY WAS REPORTED.