FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 2153517
·
Received June 21, 2011
Report
- Report Number
- 1720753-2011-08370
- Event Type
- Malfunction
- Date Received
- June 21, 2011
- Date of Event
- June 14, 2011
- Report Date
- June 21, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION AND REPLACED THE INTERCONNECT CABLE AND THE RECEPTACLE. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM'S PIN WAS BENT WHERE IT HOOKS TO THE CABLE AND CAUSED A COMMUNICATION ERROR MESSAGE AFTER A CASE. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |