VASOVIEW HEMOPRO 2 EVH SYSTEM
Report
- Report Number
- 2242352-2011-00667
- Event Type
- Malfunction
- Date Received
- June 21, 2011
- Date of Event
- May 24, 2011
- Report Date
- May 27, 2011
- Manufacturer
- MAQUET CARDIOVASCULAR, LLC
- Product Code
- GEI
- PMA / PMN Number
- K101274
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PHYSICIAN
Narratives
INVESTIGATION: VISUAL INSPECTION REVEALED THAT THERE WERE NO NON CONFORMITIES WITH THE JAWS. THERE WERE MINIMAL SIGNS OF CLINICAL USAGE AND MINIMAL EVIDENCE OF BLOOD. A PRE-CAUTERY TEST WAS PERFORMED ON THE DEVICE WITH A REFERENCE POWER SUPPLY AND CABLE. THE DEVICE PASSED THE PRE-CAUTERY TEST, IT PRODUCED STEAM AND HEAT DURING SEVERAL ACTIVATIONS. BASED UPON THE FUNCTIONAL TEST, THE REPORTED FAILURE "FAILURE TO DELIVER ENERGY" COULD NOT BE CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE PRODUCT. THERE WAS NO NONCONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. INTERNAL FILE NUMBER - (B)(4).
THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VESSEL HARVESTING PROCEDURE, THE VASOVIEW HEMOPRO 2 JAWS CAUTERIZATION STOPPED WORKING DURING SURGERY. A NEW KIT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT EFFECTS. THE PRODUCT WAS RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VASOVIEW HEMOPRO 2 EVH SYSTEM | ENDOSCOPIC VESSEL HARVESTING SYSTEM | GEI | MAQUET CARDIOVASCULAR, LLC | VH-4000 | 25030603 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |