MDR report key: 2153456 · Received June 21, 2011

Report

Report Number
3004209178-2011-04621
Event Type
Malfunction
Date Received
June 21, 2011
Date of Event
January 1, 2011
Report Date
May 24, 2011
Manufacturer
MDT PUERTO RICO OPERATIONS CO., JUNCOS
Product Code
LGW
PMA / PMN Number
P840001
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT EXPERIENCED A LOSS OF THERAPEUTIC EFFECT IN THE RIGHT LEG. THE PATIENT HAD REACHED THE MAXIMUM SETTINGS. IT WAS LATER REPORTED THAT IMPEDANCE READINGS WERE GREATER THAN 4,000 OHMS ON ALL OF THE BIPOLAR PAIRS. THE PATIENT HAD STIMULATION ON THE LEFT SIDE BUT NOT ON THE RIGHT SIDE. THE IMPEDANCE READINGS ON THE RIGHT SIDE, WITH ELECTRODES 4-7 WERE: 4/5 = 1395; 4/6 = 1395; 4/7 = 4000; 5/6 = 1395; 5/7 = 4000; 6/7 = 4000. THE PATIENT ALSO EXPERIENCED A SHOCKING OR JOLTING SENSATION WHEN THE SETTINGS WERE INCREASED DURING REPROGRAMMING. ADDITIONAL INFORMATION WAS REQUESTED BUT NOT AVAILABLE AS OF THE DATE OF THIS REPORT. A FOLLOW-UP REPORT WILL BE FILED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SYNERGY LGW MDT PUERTO RICO OPERATIONS CO., JUNCOS 7427 NA

Patients

Seq Age Sex Outcome Treatment
1 45 YR EXPLANTED:| IMPLANTED:| PROGRAMMER: MODEL 7435, LOT# NFT005431P| EXTENSION: MODEL 7495-10, LOT# NAH004597N| LEAD: MODEL 3998, LOT# L79022| EXPLANTED:| IMPLANTED:| IMPLANTED:| EXPLANTED:| EXTENSION: MODEL 7495-10, LOT# NAH004513N