FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2153430 · Received June 21, 2011

Report

Report Number
1720753-2011-08377
Event Type
Malfunction
Date Received
June 21, 2011
Date of Event
June 9, 2011
Report Date
June 21, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION AND RESEATED THE CIRCUIT BOARDS AND THE CONNECTIONS. THE VOLTAGE WAS VERIFIED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED AN INTERLOCK FAILURE ERROR MESSAGE AT BOOT-UP. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1