FDA Adverse Event Injury Summary report: N

GETINGE STEAM STERILIZER

MDR report key: 2153338 · Received June 30, 2011

Report

Report Number
1314329-2011-00003
Event Type
Injury
Date Received
June 30, 2011
Date of Event
May 19, 2011
Report Date
June 3, 2011
Manufacturer
GETINGE SOURCING LLC
Product Code
FLE
PMA / PMN Number
K020590
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

GETINGE REP WENT TO FACILITY AND INSPECTED THE STERILIZER INVOLVED IN THE INCIDENT. GETINGE REP REPORTED THAT THE FACILITY BIOMED HAD TIGHTENED LOOSE SCREWS HOLDING THE LOCKING POST TO THE INTERIOR BRACKET. USER MANUAL WAS REVIEWED - THE STEPS LISTED FOR REMOVING THE LOAD INDICATE A CHECK THAT THE CART IS SECURELY LOCKED BEFORE REMOVING THE RACK. IT IS NOT KNOWN IF THE OPERATOR PERFORMED THIS STEP PRIOR TO THE INCIDENT. THIS IS THE FIRST REPORT OF THE LOCKING NUT HARDWARE BECOMING LOOSE. FURTHER INVESTIGATION IS REQUIRED TO DETERMINE IF ANY ADDITIONAL ACTION IS NEEDED.

Description of Event or Problem · 1

USER FACILITY NOTIFIED GETINGE REP THAT AN INCIDENT HAD OCCURRED WITH THE 733HC STERILIZER DURING UNLOADING. THE LOAD CART WAS NOT PROPERLY ATTACHED TO THE LOCKING POST/TRACK ASSEMBLY LOCATED IN THE INTERIOR OF THE STERILIZER AND WHEN THE OPERATOR WENT TO PULL THE CART OUT, THE RACK BEGAN FALLING. THE OPERATOR CAUGHT THE RACK AND PUSHED IT BACK INTO THE STERILIZER. THE RACK TIPPED PINNING THE OPERATOR'S ARM BETWEEN THE RACK AND THE INTERIOR STERILIZER CHAMBER WALL. OPERATOR WAS SENT TO THE ER FOR TREATMENT TO BURNS ON HIS FOREARM. THE OPERATOR WAS TREATED AND RELEASED WITH 2ND DEGREE BURNS TO THE FOREARM. OPERATOR REPORTED BACK TO WORK WITH IN 2 DAYS. THE STERILIZER IS MAINTAINED BY THE FACILITY'S BIOMED. THE BIOMED REPORTED THAT THERE WERE LOOSE SCREWS FOUND ON THE BRACKET THAT HOLD THE LOCKING POST IN PLACE. SCREWS WERE TIGHTENED BY THE BIOMED AND NO FURTHER PROBLEMS HAVE BEEN NOTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GETINGE STEAM STERILIZER STEAM STERILIZER FLE GETINGE SOURCING LLC 733HC

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention