CONTAK RENEWAL
Report
- Report Number
- 2124215-2011-06586
- Event Type
- Malfunction
- Date Received
- July 7, 2011
- Date of Event
- March 23, 2011
- Report Date
- April 14, 2011
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- NIK
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, DEVICE INTERROGATION REVEALED THAT THE DEVICE BATTERY STATUS WAS ELECTIVE REPLACEMENT INDICATOR (ERI), WITH A BATTERY VOLTAGE OF [X.XX] VOLTS. A LONGEVITY CALCULATION CONFIRMED THAT THE DEVICE DID NOT MEET LONGEVITY ESTIMATES PROVIDED IN DEVICE LABELING. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. FINAL ANALYSIS CONCLUDED THAT THE PREMATURE BATTERY DEPLETION WAS DUE TO TWO COMPROMISED LOW-VOLTAGE CAPACITORS, WHICH RESULTED IN A HIGH CURRENT CONDITION. NO OTHER ADVERSE EVENTS HAVE BEEN REPORTED. THIS ISSUE WILL BE RE-EVALUATED IF ADDITIONAL INFORMATION IS RECEIVED.
UPON RECEIPT AT OUR (B)(4) LABORATORY, DEVICE INTERROGATION REVEALED THAT THE DEVICE BATTERY STATUS WAS ELECTIVE REPLACEMENT INDICATOR (ERI), WITH A BATTERY VOLTAGE OF 2.4 VOLTS. A LONGEVITY CALCULATION CONFIRMED THAT THE DEVICE DID NOT MEET LONGEVITY ESTIMATES PROVIDED IN DEVICE LABELING. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. FINAL ANALYSIS CONCLUDED THAT THE PREMATURE BATTERY DEPLETION WAS DUE TO TWO COMPROMISED LOW-VOLTAGE CAPACITORS, WHICH RESULTED IN A HIGH CURRENT CONDITION. NO OTHER ADVERSE EVENTS HAVE BEEN REPORTED. THIS ISSUE WILL BE RE-EVALUATED IF ADDITIONAL INFORMATION IS RECEIVED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS EXPLANTED DUE TO NORMAL ELECTIVE REPLACEMENT INDICATOR (ERI) BATTERY STATUS AND WAS REPLACED WITH ANOTHER ICD. UPON RECEIPT, INITIAL LABORATORY ANALYSIS DETERMINED THAT THIS DEVICE DID NOT MEET EXPECTED LONGEVITY PREDICTIONS AS DESCRIBED IN DEVICE LABELING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONTAK RENEWAL | IMPLANTABLE CHF GENERATOR | NIK | GUIDANT CRM CLONMEL IRELAND | H170 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | (B)(4)| (B)(4)| (B)(4)| MISMATCH| (B)(4) |