VASOVIEW HEMOPRO 2 EVH SYSTEM

MDR report key: 2153217 · Received June 17, 2011

Report

Report Number
2242352-2011-00645
Event Type
Malfunction
Date Received
June 17, 2011
Date of Event
May 24, 2011
Report Date
May 24, 2011
Manufacturer
MAQUET CARDIOVASCULAR, LLC.
Product Code
GEI
PMA / PMN Number
K101274
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION: VISUAL INSPECTION REVEALED THAT THERE WERE NO NON-CONFORMITIES WITH THE CANNULA. THERE WAS SOME EVIDENCE OF BLOOD. A FUNCTIONAL FLOW TEST WAS PERFORMED AND FLUID PASSED THROUGH THE INSUFFLATION TUBING. BASED UPON THIS, THE REPORTED COMPLAINT COULD NOT BE CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE PRODUCT. THERE WAS NO NON-CONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. INTERNAL FILE NUMBER - (B)(4).

Description of Event or Problem · 1

THE HOSP REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE VASOVIEW HEMOPRO 2 COULD NOT MAINTAIN THE TUNNEL BECAUSE THERE WAS NO INSUFFLATION. A REPLACEMENT UNIT WAS USED TO COMPLETE THE PROCEDURE. THE HOSP DID NOT REPORT ANY PT EFFECTS. THE PRODUCT WAS RETURNED.