FDA Adverse Event
Malfunction
Summary report: N
VASOVIEW HEMOPRO 2 EVH SYSTEM
MDR report key: 2153217
·
Received June 17, 2011
Report
- Report Number
- 2242352-2011-00645
- Event Type
- Malfunction
- Date Received
- June 17, 2011
- Date of Event
- May 24, 2011
- Report Date
- May 24, 2011
- Manufacturer
- MAQUET CARDIOVASCULAR, LLC.
- Product Code
- GEI
- PMA / PMN Number
- K101274
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
INVESTIGATION: VISUAL INSPECTION REVEALED THAT THERE WERE NO NON-CONFORMITIES WITH THE CANNULA. THERE WAS SOME EVIDENCE OF BLOOD. A FUNCTIONAL FLOW TEST WAS PERFORMED AND FLUID PASSED THROUGH THE INSUFFLATION TUBING. BASED UPON THIS, THE REPORTED COMPLAINT COULD NOT BE CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE PRODUCT. THERE WAS NO NON-CONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. INTERNAL FILE NUMBER - (B)(4).
Description of Event or Problem · 1
THE HOSP REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE VASOVIEW HEMOPRO 2 COULD NOT MAINTAIN THE TUNNEL BECAUSE THERE WAS NO INSUFFLATION. A REPLACEMENT UNIT WAS USED TO COMPLETE THE PROCEDURE. THE HOSP DID NOT REPORT ANY PT EFFECTS. THE PRODUCT WAS RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VASOVIEW HEMOPRO 2 EVH SYSTEM | ENDOSCOPIC VESSEL HARVESTING SYSTEM | GEI | MAQUET CARDIOVASCULAR, LLC. | VH-4000 | 25033709 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |