FDA Adverse Event Malfunction Summary report: N

HS III PROXIMAL SEAL SYTEM 4.3MM

MDR report key: 2153211 · Received June 17, 2011

Report

Report Number
2242352-2011-00637
Event Type
Malfunction
Date Received
June 17, 2011
Date of Event
May 11, 2011
Report Date
May 23, 2011
Manufacturer
MAQUET CARDIOVASCULAR, LLC.
Product Code
DXC
PMA / PMN Number
K080169
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION: VISUAL INSPECTION REVEALED THAT THE DELIVERY TUBE WAS RETURNED INSIDE THE LOADING DEVICE. THE SEAL AND THE TENSION SPRING ASSEMBLY WERE INSIDE THE LOADING DEVICE, UNDER THE WINDOW. THE GREEN SLIDE LOCK AND THE WHITE PLUNGER WERE NOT DEPRESSED ON THE DELIVERY DEVICE. THERE WAS NO EVIDENCE OF BLOOD ON THE DEVICE. BASED UPON THE RECEIVED CONDITION OF THE DEVICE, THE REPORTED COMPLAINT "SEAL WOULD NOT LOAD PROPERLY" WAS CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NON-CONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. (B)(4).

Description of Event or Problem · 1

THE HOSP REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE, THE HEARTSTRING III SEAL WOULD NOT LOAD PROPERLY. THE PER STATED "BECAUSE TENSION SPRING WAS STUCK IN DELIVERY DEVICE, PROXIMAL SEAL WAS NOT REMOVED FULLY." A NEW KIT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PT EFFECTS. THE PRODUCT WAS RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HS III PROXIMAL SEAL SYTEM 4.3MM PROXIMAL SEAL DXC MAQUET CARDIOVASCULAR, LLC. HSK-3043 25024416

Patients

Seq Age Sex Outcome Treatment
1 NA