HS III PROXIMAL SEAL SYTEM 3.8MM
Report
- Report Number
- 2242352-2011-00672
- Event Type
- Malfunction
- Date Received
- June 17, 2011
- Report Date
- May 27, 2011
- Manufacturer
- MAQUET CARDIOVASCULAR, LLC
- Product Code
- DXC
- PMA / PMN Number
- K080169
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
Narratives
INVESTIGATION: VISUAL INSPECTION REVEALED THAT THE DELIVERY TUBE WAS RETURNED INSIDE THE LOADING DEVICE. THE SEAL AND THE TENSION SPRING ASSEMBLY WERE INSIDE THE BODY OF THE LOADING DEVICE, WITH THE SEAL ROLLED. THE GREEN SLIDE LOCK AND THE WHITE PLUNGER WERE NOT DEPRESSED ON THE DELIVERY DEVICE. THERE WAS NO EVIDENCE OF BLOOD ON THE DEVICE. BASED UPON THE RECEIVED CONDITION OF THE DEVICE, THE REPORTED COMPLAINT "SEAL DID NOT LOAD PROPERLY" WAS CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NONCONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. (B)(4).
THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE, THE HEARTSTRING III SEAL DID NOT LOAD PROPERLY. A REPLACEMENT UNIT WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PATIENT EFFECTS. THE PRODUCT WAS RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HS III PROXIMAL SEAL SYTEM 3.8MM | PROXIMAL SEAL | DXC | MAQUET CARDIOVASCULAR, LLC | HSK-3038 | 25023135 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |