FDA Adverse Event Malfunction Summary report: N

COULTER LH 750 HEMATOLOGY ANALYZER

MDR report key: 2153102 · Received June 17, 2011

Report

Report Number
1061932-2011-00638
Event Type
Malfunction
Date Received
June 17, 2011
Date of Event
September 25, 2008
Report Date
September 26, 2008
Manufacturer
BECKMAN COULTER, INC.
Product Code
GKZ
PMA / PMN Number
K061574
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

ALL SAMPLES WERE FRESH DRAWS. THE INSTRUMENT WAS PERFORMING WITHIN QUALITY CONTROL SPECS. CONTROLS WERE RUN EVERY SHIFT (THREE TIMES PER DAY). FIVE C CONTROLS WERE RUN BEFORE AND AFTER THE INCIDENT AND RECOVERED WITHIN SPECS. REVIEW OF LATRON CONTROLS SUBMITTED SHOWED CONDUCTIVITY WAS HIGH OUT ON (B)(4) 2008. ON (B)(4) 2008, THE FIELD SERVICE ENGINEER (FSE) ADJUSTED THE VOLTAGE WHICH WAS FOUND TOO HIGH FOR VOLUME AND REPLACED THE FLOW CELL TO FIX THE PROBLEM WITH THE RETICULOCYTE BACKGROUNDS READING +++ AT START UP. ON (B)(4) 2008, FSE CLEANED AND ADJUSTED THE FLOW CELL AND LIGHT SCATTER SENSOR. RAW DATA FILES WERE NO LONGER AVAILABLE FOR ANALYSIS. ROOT CAUSE WAS ATTRIBUTED TO MISALIGNMENT/DIRT IN THE TRIPLE TRANSDUCER MODULE AREA. THIS REPORTABLE EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW OF COMPLAINTS CONDUCTED BETWEEN JANUARY 1, 2008 AND OCTOBER 23, 2010 FOR ADD'L REPORTABLE EVENTS. THIS MDR REPRESENTS EVENT 7 OF 9 REPORTED BY THIS CUSTOMER. THIS MDR IS RELATED TO THE FOLLOWING MDRS THAT HAVE BEEN REPORTED: MDR 1061932-2011-00632, 00633, 00634, 00635, 00636, 00637, 00639, 00640.

Description of Event or Problem · 1

CUSTOMER REPORTED ERRONEOUS ELEVATED EOSINOPHIL% AND LOW NEUTROPHIL% FOR 22 PTS TESTED BETWEEN THE DATES OF (B)(6) 2008. TESTING WAS PERFORMED USING A COULTER LH 750 HEMATOLOGY ANALYZER. THE TEST RESULTS WERE DETERMINED TO BE ERRONEOUS WHEN COMPARED TO A SECOND ANALYZER WHICH WAS CONSIDERED CORRECT. ERRONEOUS TEST RESULTS WERE REPORTED OUT OF THE LAB. NOTIFICATION TO ALL NURSES AND PHYSICIANS INVOLVED WITH THE PTS WITH INCORRECT TEST REPORTS WAS COMPLETED. THERE WERE NO REPORTS OF ADVERSE EFFECTS TO PT TREATMENT. NO REPORTS OF DEATH OR SERIOUS INJURY, AND NO AFFECT TO PT TREATMENT AS A RESULT OF THIS EVENT. THIS EVENT REPRESENTS EVENT 7 OF 9 EVENTS REPORTED BY THIS CUSTOMER FOR THE 22 PTS INVOLVED. THIS MDR IS FOR (B)(4) 2008, EVENING WORK SHIFT. AN MDR WAS SUBMITTED FOR EACH WORK SHIFT OF EACH DAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COULTER LH 750 HEMATOLOGY ANALYZER GKZ BECKMAN COULTER, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 20 YR