FDA Adverse Event Malfunction Summary report: N

STRATA

MDR report key: 2153091 · Received June 17, 2011

Report

Report Number
2021898-2011-00133
Event Type
Malfunction
Date Received
June 17, 2011
Date of Event
May 26, 2011
Report Date
May 26, 2011
Manufacturer
MEDTRONIC NEUROSURGERY
Product Code
JXG
PMA / PMN Number
K042465
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE VALVE WAS PATENT AND PASSES SIPHON TESTING. HOWEVER, IT DID NOT MEET SPECIFICATIONS FOR REFLUX OR LEAK TESTING AS THERE WERE TEARS IN THE SIDE AND TOP OF THE DELTA CHAMBER. IT IS UNK HOW OR WHEN THIS DAMAGE OCCURRED. THE DEVICE MET ALL REQUIREMENTS FOR PRESSURE/FLOW AND PRE-IMPLANTATION TESTS. THE INSTRUCTIONS FOR USE THAT ACCOMPANY THE DEVICE CAUTION THAT CARE SHOULD BE TAKEN IN HANDLING THE VALVE AS SILICONE HAS A LOW CUT AND TEAR RESISTANCE. IT IS ALSO SUGGESTS THAT THE VALVE BE PLACED IN A SURGICALLY CREATED LOOSE SUBGALEAL POCKET. THIS ALLOWS GENTLE PLACEMENT OF THE VALVE AND MINIMIZES HANDLING WITH SURGICAL TOOLS. A REVIEW OF THE MFG RECORDS SHOWED NO ANOMALIES. NO IMPACT TO THE PT WAS REPORTED.

Description of Event or Problem · 1

IT WAS REPORTED TO MEDTRONIC NEUROSURGERY THAT ALLEGEDLY A STRATA VALVE TAKEN OFF THE SHELF WAS LEAKING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STRATA JXG - SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS JXG MEDTRONIC NEUROSURGERY C76642

Patients

Seq Age Sex Outcome Treatment
1