FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 2153044 · Received June 17, 2011

Report

Report Number
8020893-2011-00264
Event Type
Malfunction
Date Received
June 17, 2011
Date of Event
June 9, 2011
Report Date
June 9, 2011
Manufacturer
COVIDIEN, FORMERLY NELLCO
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

THE SERVICE REPORT SHOWS THE CUSTOMER REPORTED THAT THE 840 VENTILATOR STOPPED CYCLING WHILE IN USE ON A PATIENT. THE PATIENT WAS NOT HARMED OR INJURED AS A RESULT OF THE EVENT. THE NELLCOR PURITAN BENNETT CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION IN THE MEMORY BUT THE CSE WAS NOT ABLE TO DUPLICATE THE PROBLEM. THE CSE INSPECTED THE DEVICE AND REPLACED THE RESPIRATORY PCB AS PRECAUTIONARY ACTION. THE UNIT PASSED EXTENDED SELF TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 840 VENTILATOR VENTILATOR CBK COVIDIEN, FORMERLY NELLCO 840

Patients

Seq Age Sex Outcome Treatment
1