FDA Adverse Event Malfunction Summary report: N

COOL PATH CATHETER 7F, 1304-CP-7-25-M-AB

MDR report key: 2153038 · Received June 17, 2011

Report

Report Number
2030404-2011-00186
Event Type
Malfunction
Date Received
June 17, 2011
Report Date
May 25, 2011
Manufacturer
ST. JUDE MEDICAL, IRVINE
Product Code
OAD
PMA / PMN Number
P060019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

VISUAL INSPECTION OF THE RETURNED CATHETER REVEALED THE LUER EXTENSION TUBE BROKE AT THE HANDLE EDGE AND SEPARATED FROM THE HANDLE. FURTHER FUNCTIONAL TESTING OF THE CATHETER COULD NOT BE PERFORMED DUE TO THIS SEPARATION. REVIEW OF THE DEVICE HISTORY RECORDS CONFIRMED THE DEVICE MET MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. THE ROOT CAUSE CLASSIFICATION IS CONSISTENT WITH SUPPLIER QUALITY. A QUALITY IMPROVEMENT PROJECT WAS INITIATED TO INVESTIGATE THE ISSUE AND IMPROVE THE PROCESS. AS A RESULT, IMPROVEMENTS TO THE TUBING QUALITY AND IMPROVEMENTS TO THE INTERNAL INSPECTION PROCESS HAVE BEEN INSTITUTED. SJM PARTNERED WITH THE VENDOR OF THE TUBING AND DETERMINED THAT THE MATERIAL WAS NOT PROCESSED WITH THE NECESSARY CONDITIONS TO ENSURE OPTIMUM TUBING QUALITY; THE PROCESS HAD BEEN IMPROVED BY THE VENDOR IN ORDER TO PREVENT POTENTIALLY DEFECTIVE TUBING FROM BEING UTILIZED DURING THE MANUFACTURING PROCESS, SJM HAS ADDITIONALLY INSTITUTED A NEW TESTING FIXTURE THAT IMPROVES OUR ABILITY TO DETECT THE NONCONFORMING TUBING.

Description of Event or Problem · 1

IT WAS REPORTED WHILE USING THE COOL PATH CATHETER FOR AN ABLATION PROCEDURE, THE IRRIGATION PORT DETACHED FROM THE PROXIMAL END OF THE HANDLE. THE PHYSICIAN EXCHANGED THE CATHETER AND COMPLETED THE PROCEDURE WITH NO CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED BUT IS NOT AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COOL PATH CATHETER 7F, 1304-CP-7-25-M-AB CARDIAC ABLATION PERCUTANEOUS CATHETER OAD ST. JUDE MEDICAL, IRVINE 83503 3208102

Patients

Seq Age Sex Outcome Treatment
1