FDA Adverse Event Injury Summary report: N

INFANT DELIVERY SYSTEM

MDR report key: 2152913 · Received July 7, 2011

Report

Report Number
9611451-2011-00411
Event Type
Injury
Date Received
July 7, 2011
Date of Event
April 8, 2011
Report Date
May 9, 2011
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BZD
PMA / PMN Number
SEE H.10
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE BC171 INFANT DELIVERY SYSTEM IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT WHICH IS SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K100011. FISHER & PAYKEL HEALTHCARE WAS FIRST NOTIFIED VIA A PHONE CALL OF THIS INCIDENT ON (B)(6) 2011. A REPORT OF THIS INCIDENT FROM THE HOSPITAL STAFF WAS NOT RECEIVED UNTIL THE (B)(6) 2011. THE COMPLAINT DEVICE HAS NOT BEEN RETURNED TO FISHER & PAYKEL HEALTHCARE FOR EVALUATION. ATTEMPTS TO OBTAIN FURTHER INFORMATION WITH REGARD TO THIS INCIDENT HAS BEEN UNSUCCESSFUL. OUR ANALYSIS IS ACCORDINGLY BASED ON THE EVENT DESCRIPTION AND OUR KNOWLEDGE OF THE PRODUCT. A LOT CHECK COULD NOT BE PERFORMED AS NO LOT NUMBER WAS PROVIDED. THE HOSPITAL HAS REPORTED THAT HOSPITAL STAFF MEMBERS HAVE NOTED AT VARIOUS STAGES THE CONDITION OF THE PATIENT'S NARES OVER A 2 WEEK PERIOD. THE HOSPITAL HAS REPORTED ALTERNATING BETWEEN MASK AND PRONGS DEPENDING ON THE CONDITION OF THE NOSE. UP UNTIL THE 12TH DAY OF THERAPY, HOSPITAL STAFF NOTED THAT THE NOSE AT TIMES WAS SLIGHTLY RED, AND A DUODERM DRESSING WAS APPLIED. LATER ON THE 12TH DAY, IT WAS NOTED THAT THE NASAL SEPTUM WAS "PINK AND HEALTHY". ON THE 14TH DAY OF THERAPY, IT WAS NOTED THAT "SKIN ON THE NOSE WAS BROKEN" AND, DESPITE USING A MASK AND THE DUODERM DRESSING, THERE WAS EVIDENCE OF FURTHER BREAKDOWN AND AS A RESULT THE PATIENT WAS REINTUBATED. A FISHER & PAYKEL HEALTHCARE REPRESENTATIVE HAS VISITED THE HOSPITAL UPON RECEIPT OF THIS COMPLAINT TO ORGANISE FURTHER TRAINING TO REITERATE USER INSTRUCTIONS. OUR USER INSTRUCTIONS ILLUSTRATE THE PROPER SET UP OF THE INTERFACES AND STATE THE FOLLOWING: "CHECK PATIENT FREQUENTLY FOR SIGNS OF REDNESS, PRESSURE SORES OR IRRITATION WHICH MAY RESULT FROM EXTENDED PRONG OR MASK USE. WE RECOMMEND HOURLY CHECKS. ALTERNATING BETWEEN MASK AND PRONGS CAN REDUCE THE RISK OF IRRITATION. ALTERNATING EVERY 4 TO 6 HOURS IS RECOMMENDED." "IF USING PRONGS: CLEAR NASAL SECRETIONS BEFORE INSERTING THE NASAL PRONGS. ENSURE THE NASAL PRONGS ARE POSITIONED AT LEAST 2MM FROM THE SEPTUM TO AVOID PRESSURE NECROSIS AND ADJUST AS NECESSARY. HOURLY CHECKS OF THE SEPTUM INTEGRITY ARE RECOMMENDED." "IF USING MASK: CONNECT TO PATIENT BY PLACING MASK AROUND NOSE. THE MASK SHOULD SIT COMFORTABLY AROUND THE PATIENT'S NOSE. IT MUST NOT OCCLUDE THE NOSTRILS OR TOUCH THE SEPTUM AND SHOULD NOT BE OVER THE LIPS OR OVER THE EYES.

Additional Manufacturer Narrative · 1

(B)(4). THE (B)(4) INFANT DELIVERY SYSTEM IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT WHICH IS SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K100011. FISHER & PAYKEL HEALTHCARE WAS FIRST NOTIFIED VIA A PHONE CALL OF THIS INCIDENT ON (B)(6) 2011. A REPORT OF THIS INCIDENT FROM THE HOSPITAL STAFF WAS NOT RECEIVED UNTIL THE (B)(6) 2011. WE ARE CURRENTLY ENDEAVOURING TO OBTAIN FURTHER INFORMATION WITH REGARD TO THIS INCIDENT. WE WILL PROVIDE A FOLLOW-UP REPORT UPON COMPLETION OF OUR INVESTIGATION.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) REPRESENTATIVE THE FOLLOWING INFORMATION INVOLVING A (B)(4) INFANT DELIVERY SYSTEM - "AN INFANT WAS PLACED ON NCPAP ON (B)(6) 2011. NOSE BECAME SORE ON (B)(6) 2011 SO PLACED ON MASK TILL THE NEXT DAY, THEN BACK ON NCPAP. ON (B)(6) 2011 NOTED NOSE SLIGHTLY RED AND BY (B)(6) 2011 SHOWED EVIDENCE OF BREAKDOWN."

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) REPRESENTATIVE THE FOLLOWING INFORMATION INVOLVING A BC171 INFANT DELIVERY SYSTEM - "AN INFANT WAS PLACED ON NCPAP ON (B)(6) 2011. NOSE BECAME SORE ON (B)(6) 2011 SO PLACED ON MASK TILL THE NEXT DAY, THEN BACK ON NCPAP. ON (B)(6) 2011, NOTED NOSE SLIGHTLY RED AND BY (B)(6) 2011 SHOWED EVIDENCE OF BREAKDOWN."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFANT DELIVERY SYSTEM BZD BZD FISHER & PAYKEL HEALTHCARE LTD BC171 NOT PROVIDED

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention