UNICEL® DXC 600I SYNCHRON® ACCESS® CLINICAL SYSTEM
Report
- Report Number
- 2050012-2011-02913
- Event Type
- Malfunction
- Date Received
- July 7, 2011
- Date of Event
- June 6, 2011
- Report Date
- June 6, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K060256
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(4) 2011 AND REPLACED BOTH SAMPLE PROBES. THE FSE ALSO REPLACED THE WASH COLLARS AND WASH COLLAR VACUUM TUBING, HOME SENSOR, THE SAMPLE WHEEL HOME SENSOR, AND THE UPPER FIBER-OPTIC HARNESS. THE FSE PERFORMED ALIGNMENT AND CLEANED THE GLUCOSE CUP. THE FSE ALSO REPLACED THE GLUCOSE SENSOR. THE FSE RETURNED AGAIN ON (B)(4) 2011 AND FOUND THAT THE GEL BUILD UP CONTINUES IN THE MC AND CC SAMPLE PROBES. THE FSE REMOVED THE SAMPLE RACK LOAD ASSEMBLY FROM THE INSTRUMENT AND FOUND THAT THE SAMPLE WHEEL MOTOR WAS RUBBING A PIECE OF GRAY PLASTIC ON THE LEFT SIDE OF THE BOX WHERE THE SAMPLE WHEEL MOTOR SITS IN. THE FSE THEN INSTALLED THE PLASTIC PROTECTOR CORRECTLY AND REPLACED THE SAMPLE WHEEL MOTOR, THE SMART MODULE THAT DRIVES THE SAMPLE WHEEL MOTOR, BOTH MC AND CC PROBES AND WASH COLLARS, BOTH MC AND CC VACUUM TUBING AND VACUUM VALVES. PER THE FSE, THE NECESSARY ALIGNMENTS WERE PERFORMED AND QC WAS RUN WITH NO ERRORS.
A CUSTOMER CONTACTED BECKMAN COULTER INC (BCI) TO REPORT GLUCOSE ERRORS AND BUILD UP COMING FROM BOTH THE CHEMISTRY CARTRIDGE (CC) AND THE MODULAR CHEMISTRY (MC) PROBES IN THE UNICEL DXC 600I SYNCHRON ACCESS CLINICAL SYSTEM. NO INJURY OR OPERATOR EXPOSURE WAS REPORTED FOR THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 600I SYNCHRON® ACCESS® CLINICAL SYSTEM | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | DXC 600I | N/A |