TANGO OPTIMO
Report
- Report Number
- 9610824-2011-00086
- Event Type
- Malfunction
- Date Received
- June 27, 2011
- Date of Event
- June 6, 2011
- Report Date
- June 27, 2011
- Manufacturer
- BIO-RED MEDICAL DIAGNOSTICS GMBH
- Product Code
- KSZ
- PMA / PMN Number
- BK080013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE CORRECT FUNCTION OF THE AFFECTED REAGENTS WERE PROVED BY TESTING THE RETENTION SAMPLES OF QUALITY CONTROL LABORATORY ON TANGO. ALL (B)(6) AND NEGATIVE REACTIONS WERE CORRECT. WE DID NOT OBSERVE ANY (B)(6) REACTIONS. A REVIEW OF THE BATCH RECORD DOCUMENTATION SHOWED NO IRREGULARITIES WHICH MIGHT HAVE AFFECTED NEGATIVELY THE QUALITY OF THE COMPLAINED LOTS.
THE CUSTOMER RAN SAMPLE #1 ((B)(6)) ON THE TANGO. THE SAMPLE IS FROM PATIENT WITH ANTI-D AND A HISTORY OF ANTI-JKA. PROTEIN LEVEL IS 5 G/DL. IN THE SAME BATCH, SAMPLE #2 ((B)(6)) WAS RUN, A KNOWN A POS WITH NO UNEXPECTED ANTIBODY. SAMPLE #2 WAS ANTIBODY SCREEN (B)(6) ON THE TANGO. CELL 1 AND 2 WERE 3+. SAMPLE #2 WAS RUN SEPARATELY AGAIN ON ORTHO GEL AND THE SCREEN WAS STILL (B)(6), SO THE CUSTOMER BELIEVES THAT NOT ONLY WAS THERE CARRY-OVER, BUT SAMPLE #2 ITSELF WAS CONTAMINATED. A REDRAW ON SAMPLE #2 DID HAVE A NEGATIVE ANTIBODY SCREEN. IN THE SAME BATCH, SAMPLE #3 ((B)(4)) WAS RUN ALSO. THIS SAMPLE WAS AN O POS BUT THE TANGO READ ALL THE FRONT TYPE AS 3+ INCLUDING THE NEG CONTROL, THE SCREEN WAS ALSO POSITIVE 3+ BOTH CELLS, WHEN IT SHOULD HAVE BEEN NEGATIVE. THEY RAN SAMPLE #3 SEPARATELY ON JUNE 9 ON THE TANGO FOR A PANEL AND ONLY CELL #4 WAS (B)(6), SHOWING A LITTLE CONTAMINATION LEFT IN THE SAMPLE ITSELF.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TANGO OPTIMO | AUTOMATED BLOOD BANK ANALYZER SYSTEM | KSZ | BIO-RED MEDICAL DIAGNOSTICS GMBH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | AHG ANTI-IGG SOLIDSCREEN II: LOT 7035091-04| EXP. 02/25/2012| BIOTESTCELL 1&2: LOT 8119011 |