FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 2152770
·
Received June 24, 2011
Report
- Report Number
- 1720753-2011-08507
- Event Type
- Malfunction
- Date Received
- June 24, 2011
- Date of Event
- June 13, 2011
- Report Date
- June 24, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE SERVICE REPRESENTATIVE REPLACED THE HARD DRIVE AND RELOADED THE SOFTWARE. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK IN SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM WOULD INTERMITTENTLY DISPLAY CORRUPT DRIVE ERROR MESSAGE. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |