FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2152748 · Received June 24, 2011

Report

Report Number
1720753-2011-08511
Event Type
Malfunction
Date Received
June 24, 2011
Date of Event
June 15, 2011
Report Date
June 24, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE REPORTED ISSUE COULD NOT BE DUPLICATED. THE POWER SUPPLY VOLTAGE WAS LOW AND WAS ADJUSTED UP TO 5.15 VOLTS. THE NODES WERE FLASHED AND THE CALIBRATION FILES WERE RELOADED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED A CHARGER ERROR MESSAGE, WOULD NOT PERFORM FLUOROSCOPY AND LOCKED UP. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900