SYMBIQ DUAL CHANNEL
Report
- Report Number
- 9615050-2011-00453
- Event Type
- Malfunction
- Date Received
- June 24, 2011
- Date of Event
- May 11, 2011
- Report Date
- May 26, 2011
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K041550
- Removal / Correction Number
- Z-1895-2010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP PASSED TESTING BY APPROPRIATELY ALARMING WHEN AIR WAS PRESENT IN THE TUBING DISTAL TO THE PUMP. THIS PUMP HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THE HISTORY WAS DOWNLOADED AT THE SERVICE CENTER. A REVIEW OF THE HISTORY INDICATED THAT ON (B)(6) 2011 AT 0210, CH A WAS PROGRAMMED IN THE BASIC MODE, USING THE DRUG LIBRARY TO DELIVER PROPOFOL 100ML, DOSE OF 40MCG/KG/MIN, AT A RATE OF 21.6ML/HR, A VTBI (VOLUME TO BE INFUSED) 3ML, FOR A DURATION OF 9 MINS, AIR IN LINE SETTING 500MCL, AND THE INFUSION WAS STARTED. BETWEEN 0210 AND 1728, THE DOSE, RATE, VTBI (LARGEST VOLUME WAS 95ML) AND DURATION WERE CHANGED 12 TIMES. CH A IS PLACED IN STANDBY MODE FROM 0430 TO 0827, WHEN THE BASIC PROGRAM IS RESUMED. AT 0827, THE INFUSION WAS STOPPED, THE CASSETTE EJECTED, CHECK CASSETTE ALARM, CASSETTE LOADED, CHECK CASSETTE ALARM CLEARED, INFUSION RESUMED, STOPPED, CHECK FLOWSTOP ALARM, ALARM SILENCED, CASSETTE EJECTED, CHECK FLOWSTOP ALARM CLEARED, CASSETTE LOADED AND INFUSION RESUMED. AT 1134, THE BASIC PROGRAM IS STOPPED, CASSETTE EJECTED, CHECK CASSETTE ALARM, CASSETTE LOADED, CHECK CASSETTE ALARM CLEARED, AND INFUSION IS RESUMED. AT 1206, THE INFUSION IS STOPPED, CHECK FLOWSTOP ALARM, ALARM SILENCED, CASSETTE EJECTED, CHECK FLOWSTOP ALARM CLEARED, CHECK CASSETTE ALARM, CASSETTE LOADED, CHECK CASSETTE ALARM CLEARED AND INFUSION RESUMED AT 10.8ML/HR. AT 1222, THE RATE WAS CHANGED TO 16.2ML/HR. AT 1346, AND END OF INFUSION ALARM OCCURS AND KVO DELIVERY BEGAN. AT 1353, END OF INFUSION ALARM WAS CLEARED, STOP KVO DELIVERY, THE VTBI WAS REPROGRAMMED FOR 2ML, FOR A DURATION OF 8 MIN, AND INFUSION WAS STARTED. AT 1401, END OF INFUSION ALARM AS OCCURRED, KVO DELIVERY BEGAN, STOP KVO DELIVERY, REPROGRAMMED A VTBI OF 95ML, FOR A DURATION OF 5 HOURS, 52 MIN AND INFUSION WAS STARTED. AT 1401, THE INFUSION WAS STOPPED, CASSETTE EJECTED, CHECK FLOWSTOP ALARM, ALARM SILENCED, CASSETTE EJECTED, AND CHECK FLOWSTOP ALARM CLEARED, CHECK CASSETTE ALARM, CASSETTE LOADED, CHECK CASSETTE ALARM CLEARED AND INFUSION WAS STARTED. AT 1404, THE INFUSION IS STOPPED, CH A IS PLACED ON STANDBY MODE FOR 30 SECONDS AND INFUSION IS STARTED. BETWEEN 1407 AND THE NEW DAY STAMP ON (B)(6) 2011 AT 1728, THERE ARE TWO CHANGES IN THE DOSE, RATE, VTBI AND DURATION. AT 1802, A USING BATTERY ALARM OCCURRED. AT 1820, USING AC ALARM OCCURRED. A REVIEW OF THE HISTORY FOUND THE PUMP DELIVERED AS PROGRAMMED. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
THE CUSTOMER CONTACT REPORTED AIR IN THE TUBING DISTAL TO THE PUMP WITH NO PUMP ALARM. AT AN UNSPECIFIED TIME, THE PUMP WAS PROGRAMMED TO DELIVER AN UNSPECIFIED CONCENTRATION OF PROPOFOL, WITH A 100ML VTBI (VOLUME TO BE INFUSED) AND THE DELIVERY WAS STARTED. THE CUSTOMER CONTACT INDICATED THAT THE CONTAINER SIZE WAS 100ML. NO FURTHER PROGRAMMING PARAMETERS WERE PROVIDED. AFTER AN UNSPECIFIED LENGTH OF TIME, THE PUMP ALARMED FOR END OF INFUSION. AT THAT TIME, THE NURSE REPORTED CONTAINER WAS EMPTY AND AIR WAS NOTED 6-8 INCHES PAST CASSETTE. IT WAS REPORTED NO AIR REACHED THE PATIENT. THE PUMP WAS REMOVED FROM CLINICAL USE. THERAPY WAS RESUMED USING A REPLACEMENT PUMP. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS AND NO REPORTED DELAY OF THERAPY CRITICAL TO THIS PATIENT. NO MEDICAL INTERVENTIONS WERE REQUIRED. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SYMBIQ DUAL CHANNEL | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | SYMBIQ TUBING SET: LIST #16005, LOT #UNK |