FDA Adverse Event Malfunction Summary report: N

9600

MDR report key: 2152633 · Received June 22, 2011

Report

Report Number
1720753-2011-08402
Event Type
Malfunction
Date Received
June 22, 2011
Date of Event
June 2, 2011
Report Date
June 22, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE SERVICE REPRESENTATIVE REPLACED THE SRAM, AND THE TECH PROCESSOR PRINTED CIRCUIT BOARDS. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING A INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT BOOT UP. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9600 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9600

Patients

Seq Age Sex Outcome Treatment
1