FDA Adverse Event Malfunction Summary report: N

6600

MDR report key: 2152631 · Received June 27, 2011

Report

Report Number
1720753-2011-08529
Event Type
Malfunction
Date Received
June 27, 2011
Date of Event
April 18, 2011
Report Date
June 27, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE ROM CHIP U25 ON THE IMAGE PROCESSOR WAS REPLACED, AND THE HARD DISK DRIVE AND THE MOTHERBOARD WERE REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM LOST DATA. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 6600 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 6600

Patients

Seq Age Sex Outcome Treatment
1