FDA Adverse Event Malfunction Summary report: N

8800

MDR report key: 2152618 · Received June 22, 2011

Report

Report Number
9617766-2011-01275
Event Type
Malfunction
Date Received
June 22, 2011
Date of Event
June 9, 2011
Report Date
June 22, 2011
Manufacturer
GE MEDICAL SYSTEMS (INDIA) PRIVATE LTD.
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE X-RAY TUBE WAS REPAIRED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED A GENERATOR ERROR MESSAGE. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 8800 FLLUOROSCOPIC X-RAY JAA GE MEDICAL SYSTEMS (INDIA) PRIVATE LTD. 8800

Patients

Seq Age Sex Outcome Treatment
1