MDR report key: 2152596 · Received June 22, 2011

Report

Report Number
1720753-2011-08396
Event Type
Malfunction
Date Received
June 22, 2011
Date of Event
June 2, 2011
Report Date
June 22, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE SERVICE REP REPLACED AND FORMATTED THE CINE DRIVE, RELOADED THE SOFTWARE, FLASHED THE NODES, AND REBUILT THE SYSTEM FILES. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK IN SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WOULD DISPLAY A RED COLOR ON THE RIGHT MONITOR AND THE DISC IS NOT AVAILABLE ERROR MESSAGE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800