FDA Adverse Event Malfunction Summary report: N

9600

MDR report key: 2152592 · Received June 27, 2011

Report

Report Number
1720753-2011-08542
Event Type
Malfunction
Date Received
June 27, 2011
Date of Event
June 16, 2011
Report Date
June 27, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE SERVICE REPRESENTATIVE REPLACED THE TECHNICAL PROCESSOR PRINTED CIRCUIT BOARD, AND THE SRAM CARD, AND RECONFIGURED THE OBTUSER FILE. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK IN SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT MOVE UP OR DOWN. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9600 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9600

Patients

Seq Age Sex Outcome Treatment
1