TERUMO VIRTUOSAPH VEIN HARVESTING SYSTEM
Report
- Report Number
- 1124841-2011-00276
- Event Type
- Malfunction
- Date Received
- June 27, 2011
- Date of Event
- June 7, 2011
- Report Date
- June 7, 2011
- Manufacturer
- TERUMO CARDIOVASCULAR SYSTEMS CORP.
- Product Code
- GCJ
- PMA / PMN Number
- K083194
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
TERUMO DID RECEIVE THE ACTUAL DEVICE; VISUAL INSPECTION NOTED MINOR SCRATCHES, DENTS ON THE ROD AND EYEPIECE OF THE ENDOSCOPE. THIS TYPE OF DAMAGE IS OFTEN ATTRIBUTED TO HANDLING ISSUES. THE DEVICE IFU STATES THAT THE ENDOSCOPE MUST BE HANDLED WITH EXTREME CARE TO PREVENT DAMAGE TO THE DEVICE. (B)(4). THE DEVICE IFU STATES TO CHECK THE IMAGE QUALITY BEFORE AND AFTER EACH USE TO CHECK FOR SIGNS OF DAMAGE. IF IMAGE IS UNACCEPTABLE, CONTACT TERUMO CVS CUSTOMER SERVICE FOR ASSISTANCE. ALL AVAILABLE INFO HAS BEEN PLACED ON FILE IN QUALITY MGMT FOR APPROPRIATE TRACKING, TRENDING AND FOLLOW UP.
THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR SYSTEMS THAT PRIOR TO ENDOSCOPIC VEIN HARVESTING, DURING SETUP, THE ENDOSCOPE WAS BLURRY. THE PRODUCT WAS CHANGED OUT. THERE WAS NO PT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TERUMO VIRTUOSAPH VEIN HARVESTING SYSTEM | ENDOSCOPE | GCJ | TERUMO CARDIOVASCULAR SYSTEMS CORP. | MCENDO550R | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |