FDA Adverse Event Malfunction Summary report: N

7900

MDR report key: 2152560 · Received June 22, 2011

Report

Report Number
9680959-2011-01479
Event Type
Malfunction
Date Received
June 22, 2011
Date of Event
June 13, 2011
Report Date
June 22, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS GMBH
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE REPORTED ISSUE COULD NOT BE DUPLICATED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WAS NOT WORKING, DISPLAYED MULTIPLE ERRORS, WOULD NOT PERFORM FLUOROSCOPY, AND THERE WERE POWER SURGES. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 7900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS GMBH 7900

Patients

Seq Age Sex Outcome Treatment
1