FDA Adverse Event
Malfunction
Summary report: N
SYMBIQ SINGLE CHANNEL
MDR report key: 2152559
·
Received June 27, 2011
Report
- Report Number
- 9615050-2011-00461
- Event Type
- Malfunction
- Date Received
- June 27, 2011
- Date of Event
- February 2, 2011
- Report Date
- February 2, 2011
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K041550
- Removal / Correction Number
- Z-1596-2011
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RECEIVED ON (B)(4) 2011. DURING TESTING CHARRING WAS FOUND ON THE NEUTRAL AND LINE VOLTAGE BLADES AND GROUND PRONG OF THE AC POWER CORD PLUG. THE DEVICE PASSED THE ELECTRICAL SAFETY TEST. THE CAUSE OF THE CHARRING ON THE AC POWER CORD PLUG COULD NOT BE DETERMINED. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THE EVENT THAT IS BEING REPORTED WAS NOTED DURING VERIFICATION TESTING; THEREFORE, USER FACILITY EVENT CODES ARE NOT APPLICABLE IN THIS CASE. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
Description of Event or Problem · 1
DURING VERIFICATION TESTING AT THE SERVICE CENTER, THE MALE END OF THE AC POWER CORD PLUG WAS FOUND TO BE CHARRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SYMBIQ SINGLE CHANNEL | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |