FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2152520 · Received June 27, 2011

Report

Report Number
1720753-2011-08561
Event Type
Malfunction
Date Received
June 27, 2011
Date of Event
June 13, 2011
Report Date
June 27, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. A CLEAN LOAD OF SOFTWARE WAS PERFORMED AND THE SOFTWARE PARAMETERS WERE RELOADED. THE WIRING HARNESS IN THE WORKSTATION WAS ALSO REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WOULD INTERMITTENTLY NOT SAVE IMAGES. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1