FDA Adverse Event Malfunction Summary report: N

HS II PROXIMAL SEAL SYSTEM 4.3MM

MDR report key: 2152429 · Received June 24, 2011

Report

Report Number
2242352-2011-00742
Event Type
Malfunction
Date Received
June 24, 2011
Date of Event
May 18, 2011
Report Date
June 2, 2011
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
DXC
PMA / PMN Number
K080169
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION: VISUAL INSPECTION REVEALED THAT THE DELIVERY TUBE WAS RETURNED SEPARATE FROM THE LOADING DEVICE. THE SEAL AND THE TENSION SPRING ASSEMBLY WERE INSIDE THE BODY OF THE LOADING DEVICE. THE GREEN SLIDE LOCK AND THE WHITE PLUNGER WERE DEPRESSED ON THE DELIVERY DEVICE. THERE WAS NO EVIDENCE OF BLOOD ON THE DEVICE. BASED UPON THE RECEIVED CONDITION OF THE DEVICE, THE REPORTED COMPLAINT "SEAL GOT STUCK IN THE DEPLOYMENT DEVICE" WAS CONFIRMED AS A "FAILURE TO LOAD" ISSUE. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NONCONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE, THE HEARTSTRING III SEAL GOT STUCK IN THE DEPLOYMENT DEVICE. A REPLACEMENT UNIT WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PATIENT EFFECTS. THE PRODUCT WAS RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HS II PROXIMAL SEAL SYSTEM 4.3MM PROXIMAL SEAL DXC MAQUET CARDIOVASCULAR, LLC HSK-3043 25032332

Patients

Seq Age Sex Outcome Treatment
1 NA