FDA Adverse Event Malfunction Summary report: N

PUMP IN STYLE ADVANCED

MDR report key: 2152407 · Received June 28, 2011

Report

Report Number
1419937-2011-00137
Event Type
Malfunction
Date Received
June 28, 2011
Date of Event
May 15, 2011
Report Date
May 31, 2011
Manufacturer
MEDELA, INC.
Product Code
HGX
PMA / PMN Number
K031614
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

A REPLACEMENT POWER ADAPTER WAS SENT TO THE CUSTOMER. IN FOLLOW UP WITH THE CUSTOMER, SHE INDICATED THAT SHE DID NOT SEE FIRE OR SMOKE AND THAT THERE IS NO BREACH IN THE TRANSFORMER HOUSING BUT SHE DID SEE SPARKING AND EXPOSED WIRES AT THE STRAIN RELIEF. SHE ALSO INDICATED THAT NO INJURIES WERE SUSTAINED AS A RESULT OF THE ISSUE. IN SUMMARY, A BREACH IN THE INSULATION ON THE DC OUTPUT CORD AT THE STRAIN RELIEF WAS PRESENT, EXPOSING WIRES WHICH COULD CAUSE SPARKING. AS A RESULT OF CAPA (B)(4) WHICH WAS INITIATED FOR PUMP IN STYLE TRANSFORMER OVERHEATING/MELTING ISSUES, A FIELD ACTION SAFETY NOTIFICATION WAS INITIATED ON 04/04/2011. ACTIVITIES RELATED TO THIS NOTIFICATION ARE ON-GOING. EVAL SUMMARY: A VISUAL EXAMINATION OF THE POWER SUPPLY REVEALED NO DEFORMATION ON THE TRANSFORMER HOUSING AND NO HOLES OR OPENINGS. A VISUAL EXAMINATION OF THE CORD REVEALED A LOT OF TWISTING AND A BREACH IN THE INSULATION AT THE STRAIN RELIEF, EXPOSING WIRES. THE POWER SUPPLY WAS PLUGGED IN AND THE VOLTAGE ACROSS THE DC PLUG WAS MEASURED AT 12.5 VDC, WHICH IS NORMAL AND INDICATES THAT THE POWER SUPPLY IS PRODUCING THE CORRECT VOLTAGE. RESISTANCE ACROSS THE DC PLUG WAS MEASURED AS OPEN, WHICH INDICATES THAT THERE IS NOT A SHORT IN THE DC CORD. SMOKE, FIRE AND SPARKING WERE NOT OBSERVED WHILE THE POWER SUPPLY WAS PLUGGED IN. THE RESISTANCE ACROSS THE AC BLADES MEASURED AT 63.8 OHMS, WHICH IS NORMAL AND INDICATES THAT THE THERMAL FUSE DID NOT BLOW.

Description of Event or Problem · 1

THE CUSTOMER REPORTED TO CUSTOMER SERVICE THAT THE POWER SUPPLY FOR HER PUMP HAS EXPOSED WIRES AND SPARKS EVERY TIME SHE USES IT. NO INJURIES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP IN STYLE ADVANCED HGX MEDELA, INC. UNK

Patients

Seq Age Sex Outcome Treatment
1