HEARTSTRING PROXIMAL SEAL, 4.5 MM
Report
- Report Number
- 2242352-2011-00677
- Event Type
- Malfunction
- Date Received
- June 24, 2011
- Date of Event
- March 11, 2011
- Report Date
- May 30, 2011
- Manufacturer
- MAQUET CARDIOVASCULAR, LLC
- Product Code
- DXC
- PMA / PMN Number
- K022880
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
INVESTIGATION: VISUAL INSPECTION REVEALED THAT THE TENSION SPRING ASSEMBLY WAS INSIDE THE DELIVERY TUBE AND THE SEAL WAS EXTENDED OUTSIDE THE TUBE. THE SEAL HAD CRACKED ALONG THE FOURTH RING FROM THE EDGE. THE WHITE COLLAR WAS NOT PRESENT; THE PLUNGER HAD BEEN PARTIALLY DEPRESSED. THE DEVICE WAS BLOODY. BASED UPON THE RECEIVED CONDITION OF THE DEVICE, THE REPORTED COMPLAINT "SEAL CRACKED DURING USE" WAS CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NONCONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. INTERNAL FILE NUMBER - (B)(4).
THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE, THE HEARTSTRING II SEAL CRACKED DURING USE. A REPLACEMENT UNIT WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PATIENT EFFECTS. THE PRODUCT WAS RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HEARTSTRING PROXIMAL SEAL, 4.5 MM | PROXIMAL SEAL | DXC | MAQUET CARDIOVASCULAR, LLC | HS-1045 | 25023404 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |