FDA Adverse Event Malfunction Summary report: N

HEARTSTRING PROXIMAL SEAL, 4.5 MM

MDR report key: 2152392 · Received June 24, 2011

Report

Report Number
2242352-2011-00677
Event Type
Malfunction
Date Received
June 24, 2011
Date of Event
March 11, 2011
Report Date
May 30, 2011
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
DXC
PMA / PMN Number
K022880
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION: VISUAL INSPECTION REVEALED THAT THE TENSION SPRING ASSEMBLY WAS INSIDE THE DELIVERY TUBE AND THE SEAL WAS EXTENDED OUTSIDE THE TUBE. THE SEAL HAD CRACKED ALONG THE FOURTH RING FROM THE EDGE. THE WHITE COLLAR WAS NOT PRESENT; THE PLUNGER HAD BEEN PARTIALLY DEPRESSED. THE DEVICE WAS BLOODY. BASED UPON THE RECEIVED CONDITION OF THE DEVICE, THE REPORTED COMPLAINT "SEAL CRACKED DURING USE" WAS CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NONCONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. INTERNAL FILE NUMBER - (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE, THE HEARTSTRING II SEAL CRACKED DURING USE. A REPLACEMENT UNIT WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PATIENT EFFECTS. THE PRODUCT WAS RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HEARTSTRING PROXIMAL SEAL, 4.5 MM PROXIMAL SEAL DXC MAQUET CARDIOVASCULAR, LLC HS-1045 25023404

Patients

Seq Age Sex Outcome Treatment
1 NA