HS III PROXIMAL SEAL SYTEM 4.3MM
Report
- Report Number
- 2242352-2011-00746
- Event Type
- Malfunction
- Date Received
- June 24, 2011
- Date of Event
- June 2, 2011
- Report Date
- June 2, 2011
- Manufacturer
- MAQUET CARDIOVASCULAR, LLC
- Product Code
- DXC
- PMA / PMN Number
- K080169
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- PHYSICIAN
Narratives
INVESTIGATION: VISUAL INSPECTION REVEALED THAT THE DELIVERY TUBE WAS RETURNED SEPARATE FROM THE LOADING DEVICE. THE SEAL AND THE TENSION SPRING ASSEMBLY WERE INSIDE THE LOADING DEVICE, UNDER THE WINDOW. THE GREEN SLIDE LOCK AND THE WHITE PLUNGER WERE NOT DEPRESSED ON THE DELIVERY DEVICE. THERE WAS NO EVIDENCE OF BLOOD ON THE DEVICE. BASED UPON THE RECEIVED CONDITION OF THE DEVICE, THE REPORTED COMPLAINT "SEAL WOULD NOT LOAD" WAS CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NONCONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. INTERNAL FILE NUMBER - (B)(4).
THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE, THE HEARTSTRING III SEAL WOULD NOT LOAD, IT GOT STUCK IN THE LOADER. A NEW KIT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT EFFECTS. THE PRODUCT WAS RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HS III PROXIMAL SEAL SYTEM 4.3MM | PROXIMAL SEAL | DXC | MAQUET CARDIOVASCULAR, LLC | HSK-3043 | 25029998 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |