VASOVIEW HEMOPRO EVH SYSTEM
MDR report key: 2152371
·
Received June 24, 2011
Report
- Report Number
- 2242352-2011-00696
- Event Type
- Malfunction
- Date Received
- June 24, 2011
- Date of Event
- May 27, 2011
- Report Date
- June 1, 2011
- Manufacturer
- MAQUET CARDIOVASCULAR, LLC
- Product Code
- GEI
- PMA / PMN Number
- K052274
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
INVESTIGATION: VISUAL INSPECTION REVEALED THAT THE TIP OF THE COLD JAW SILICONE BOOT WAS DETACHED AND MISSING. THE JAWS WERE BURNT. THERE WAS SOME EVIDENCE OF BLOOD. BASED UPON THE VISUAL OBSERVATIONS, THE REPORTED COMPLAINT FOR "JAW BOOT BROKE OFF" WAS CONFIRMED. THE LOT NUMBER COULD NOT BE OBTAINED SO A LOT HISTORY RECORD REVIEW COULD NOT BE PERFORMED. (B)(4).
Description of Event or Problem · 1
THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VESSEL HARVESTING PROCEDURE, THE VASOVIEW HEMOPRO DEVICE HAD THE JAWS FRACTURE INSIDE THE PATIENT'S LEG. IT WAS NOTICED TO BE CRACKED BEFORE IT ACTUALLY BROKE OFF. A NEW KIT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT EFFECTS. THE PRODUCT WAS RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VASOVIEW HEMOPRO EVH SYSTEM | ENDOSCOPIC VESSEL HARVESTING SYSTEM | GEI | MAQUET CARDIOVASCULAR, LLC | VH-3000 | NI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |