UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM
Report
- Report Number
- 2122870-2011-02029
- Event Type
- Malfunction
- Date Received
- June 24, 2011
- Date of Event
- May 20, 2008
- Report Date
- May 20, 2008
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- PMA / PMN Number
- K023764
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
FIELD SERVICE ENGINEER (FSE) WAS ONSITE (B)(4) 2008 TO (B)(4) 2008 TO INVESTIGATE THE EVENT. THE FSE INDICATED THE CUSTOMER PERFORMED A PRECISION RUN ON A TNI AND CK-MB WITH ALL THREE LEVELS OF CONTROLS AND RESULTS WERE WITHIN SPECIFICATION. THE FSE CHECKED THE MAINTENANCE RECORDS WHICH INDICATED THAT ASPIRATE PROBE/DISPENSE PROBE MAINTENANCE HAD NOT BEEN DONE SINCE (B)(4) 2008. THE FSE CHECKED SYSTEM CHECK RESULTS AND THEY WERE WITHIN SPECIFICATIONS. THE FSE PERFORMED HIGH SENSITIVITY SYSTEM CHECK WITH THE FOAM PORTION OF THE TEST FAILING. FURTHER INVESTIGATION BY THE FSE INDICATED THAT THE INSTRUMENT IS DUE FOR A PREVENTATIVE MAINTENANCE (PM). THE PM WAS COMPLETED BY THE FSE. THE FSE ADJUSTED AND REPAIRED PIPETTOR 4 ULTRASONIC'S, REPLACED CLOGGED AIR FILTER AND REPAIRED A LEAK IN THE LEAK IN THE QUICK CONNECTING FITTING FOR PIPETTOR 1. THE FSE PERFORMED A SYSTEM CHECK AND HIGH SENSITIVITY SYSTEM CHECK AFTER THE PM AND BOTH TESTS PASSED WITHIN SPECIFICATIONS. THE FSE VERIFIED REPAIRS PER ESTABLISHED PROCEDURES AND RESULTS MEET PUBLISHED PERFORMANCE SPECIFICATIONS. ALTHOUGH THE FSE ADDRESSED HARDWARE ISSUES, A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED FOR THIS EVENT. THIS IS TWO OF TWO SEPARATE MDR REPORTS RELATED TO TWO PT EVENTS ASSOCIATED WITH A SINGLE MALFUNCTION REPORT. REFERENCE MDR NUMBER 2122870-2011-02028 FOR ALL RELATED EVENTS. THIS REPORTABLE EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW OF COMPLAINTS CONDUCTED BETWEEN (B)(4) 2008 THRU (B)(4) 2010 FOR ADD'L REPORTABLE EVENTS.
THE CUSTOMER CONTACTED BECKMAN COULTER, INC (BCI) IN REGARDS TO ERRONEOUS ACCUTNI (TROPONIN I) RESULT GENERATED ON A UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM. THE CUSTOMER RE-RAN THE SAMPLES AND THE RESULTS WERE IN THE NORMAL REFERENCE RANGE. THE INITIAL ERRONEOUS RESULTS WERE REPORTED OUTSIDE THE LABORATORY. IT IS NOT KNOWN IF THERE WAS ANY CHANGE TO THE PT TREATMENT. THERE IS NO INDICATION OF AN ADVERSE EVENT OR INDICATION OF ANY MEDICAL INTERVENTION TO PREVENT SERIOUS INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM | JJE | BECKMAN COULTER, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | ACCESS ACCUTNI |