FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2152291 · Received July 7, 2011

Report

Report Number
2531779-2011-04751
Event Type
Malfunction
Date Received
July 7, 2011
Report Date
June 8, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 08/12/2011 - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/15/2011 WITH THE FOLLOWING FINDINGS: THE PUMP BLACK BOX HISTORY INDICATED THAT THERE WAS A REPLACE BATTERY ALARM AT 5:20 AM. THE PUMP THEN ENTERED A REBOOT CYCLE DUE TO A DEPLETED BATTERY. DURING INVESTIGATION, THE PUMP BOOTED TO THE "VERIFY" SCREEN WITH FUNCTIONAL ALARMS. THE PUMP WAS EXERCISED FOR 24 HOURS WITHOUT ANY INTERRUPTIONS.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2011, THE LAY-USER/PATIENT CONTACTED ANIMAS ALLEGING THAT A PUMP HAD NO POWER. THE PATIENT DID NOT ALLEGE ANY HARM OR INJURY BECAUSE OF THE REPORTED ISSUE. BASED ON THE INFORMATION PROVIDED, THIS COMPLAINT IS BEING REPORTED DUE TO THE FOLLOWING CONCLUSION: THE PATIENT CLAIMED THAT THE PUMP WOULD NOT POWER ON. THERE IS NO EVIDENCE THAT THE PUMP WAS WORKING IMPROPERLY OR THAT THE DEVICE CONTRIBUTED TO A SERIOUS INJURY. THE PATIENT DID NOT REPORT ANY BLOOD GLUCOSE READINGS, SYMPTOMS, OR TREATMENT THAT MEET ANIMAS' CRITERIA FOR A SERIOUS INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM 2020 INSULIN INFUSION PUMP

Patients

Seq Age Sex Outcome Treatment
1 19 YR