COMPACT LOW IMPEDANCE LEAD
Report
- Report Number
- 6000153-2011-04712
- Event Type
- Malfunction
- Date Received
- June 23, 2011
- Date of Event
- May 25, 2011
- Report Date
- May 25, 2011
- Manufacturer
- MPROC, VILLALBA
- Product Code
- LGW
- PMA / PMN Number
- P840001
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
(B)(4). TESTING OF THE LEADS (MODEL 3778, (B)(4)) REVEALED NO ANOMALIES. THE LEADS HAD ACCEPTABLE CONTINUITY AND NO SHORTS. THE RETURNED STYLET AND A KNOWN GOOD STYLET COULD BE INSERTED AND REMOVED WITH NO ISSUES. THE LEADS WERE FUNCTIONALLY OKAY. TESTING OF THE STYLET (MODEL STYLET/STIM, SERIAL NUMBER UNKNOWN) REVEALED NO SIGNIFICANT ANOMALIES. THE STYLET WIRE WAS BENT 14 CM FROM THE PROXIMAL END.
THE HEALTH CARE PROFESSIONAL WAS UNABLE TO INSERT ANY STYLETS INTO TWO LEADS DURING IMPLANT. THE STYLETS WERE CLEANED, BUT STILL COULD NOT BE INSERTED. TWO NEW LEADS WERE USED TO COMPLETE THE IMPLANT. THE PATIENT RECOVERED WITHOUT SEQUELA. A FOLLOW-UP REPORT WILL BE SENT IF ADDITIONAL INFORMATION BECOMES AVAILABLE. SEE ALSO MANUFACTURER'S REPORT #6000153-2011-04710.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COMPACT LOW IMPEDANCE LEAD | LGW | MPROC, VILLALBA | 3778 | V11026443 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | IMPLANTED:| LEAD: MODEL 3778, LOT# V639156031| EXPLANTED:| STIM ACCESSORY: MODEL STYLET/STIM, LOT# UNKNOWN |