INFINITI VISION SYSTEM OZIL
Report
- Report Number
- 2028159-2011-00698
- Event Type
- Malfunction
- Date Received
- June 23, 2011
- Date of Event
- May 12, 2011
- Report Date
- May 24, 2011
- Manufacturer
- ALCON - IRVINE TECHNOLOGY CENTER
- Product Code
- HQC
- PMA / PMN Number
- K082845
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- OTHER
Narratives
THE COMPANY REPRESENTATIVE EXAMINED THE SYSTEM AND COULD NOT DUPLICATE THE CUSTOMER REPORTED EVENT. THE SYSTEM EVENT LOG WAS REVIEWED AND NO SYSTEM MESSAGES WERE FOUND ASSOCIATED WITH THE CUSTOMER REPORTED EVENT. THE COMPANY REPRESENTATIVE TESTED THE CUSTOMER'S HANDPIECES AND ALL PERFORMED TO SPECIFICATIONS. PREVENTIVE MAINTENANCE WAS PERFORMED. THE SYSTEM WAS THEN TESTED AND MET ALL PRODUCT SPECIFICATIONS. THERE WAS NO SAMPLE RETURNED FOR EVALUATION AND NO ADDITIONAL INFORMATION PROVIDED RELATED TO THIS EVENT. FOR THIS REASON, STEPS COULD NOT BE TAKEN TO REPLICATE OR CONFIRM THE REPORTED EVENT. A REVIEW OF COMPLAINTS FOR THE LAST 24 MONTHS DID INDICATE TWO ADDITIONAL SIMILAR REPORTS FOR THIS SYSTEM. THE ROOT CAUSE IS UNKNOWN AT THIS TIME. (B)(4).
A HEALTHCARE PROFESSIONAL REPORTED LOW PHACO POWER WAS EXPERIENCED DURING A CATARACT EXTRACTION PROCEDURE. THE CASE TOOK LONGER TO PERFORM, HOWEVER, THERE WAS NO DELAY OR HARM TO THE PATIENT. THIS IS THE SECOND OF THREE MEDICAL DEVICE REPORTS BEING FILED FOR THIS FACILITY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INFINITI VISION SYSTEM OZIL | PHACOFRAGMENTATION SYSTEM | HQC | ALCON - IRVINE TECHNOLOGY CENTER | INFINITI OZIL | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR |