FDA Adverse Event
Malfunction
Summary report: N
INFINITI VISION SYSTEM OZIL
MDR report key: 2152245
·
Received June 23, 2011
Report
- Report Number
- 2028159-2011-00713
- Event Type
- Malfunction
- Date Received
- June 23, 2011
- Date of Event
- May 24, 2011
- Report Date
- May 24, 2011
- Manufacturer
- ALCON - IRVINE TECHNOLOGY CENTER
- Product Code
- HQC
- PMA / PMN Number
- K082845
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
THE COMPANY REPRESENTATIVE EXAMINED THE SYSTEM AND REPLACED THE HOST CPU. PREVENTIVE MAINTENANCE WAS PERFORMED. THE SYSTEM WAS THEN TESTED AND MET ALL PRODUCT SPECIFICATIONS. INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. (B)(4).
Description of Event or Problem · 1
A NURSE REPORTED DURING A PROCEDURE THE DISPLAY SCREEN WENT BLANK AND THE IV POLE WOULD NOT LOWER. ANOTHER UNIT WAS USED TO COMPLETE THE CASE. THERE WAS NO HARM OR INJURY TO THE PATIENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INFINITI VISION SYSTEM OZIL | PHACOFRAGMENTATION SYSTEM | HQC | ALCON - IRVINE TECHNOLOGY CENTER | INFINITI OZIL | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |